Literature DB >> 1781532

Continuous direct and indirect blood pressure measurement (Finapres) in the critically ill.

I K Farquhar1.   

Abstract

Continuous noninvasive blood pressure measurement is of great potential use in the critically ill. This study was designed to find out whether measurements of blood pressure by a Finapres accurately represented intra-arterial pressure. Comparisons were made between readings from a radial artery line and from a Finapres finger cuff in 10 critically ill patients. There was an unpredictable but usually stable difference between measurements from the two sources; this difference varied in magnitude and direction in individual patients. At present the Finapres cannot be recommended as a substitute for intra-arterial blood pressure measurement in the critically ill.

Entities:  

Mesh:

Year:  1991        PMID: 1781532     DOI: 10.1111/j.1365-2044.1991.tb09922.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  2 in total

1.  The normal response to prolonged passive head up tilt testing.

Authors:  M E Petersen; T R Williams; C Gordon; R Chamberlain-Webber; R Sutton
Journal:  Heart       Date:  2000-11       Impact factor: 5.994

Review 2.  Sepsis: A Review of Advances in Management.

Authors:  Jordi Rello; Francisco Valenzuela-Sánchez; Maria Ruiz-Rodriguez; Silvia Moyano
Journal:  Adv Ther       Date:  2017-10-11       Impact factor: 3.845

  2 in total