Literature DB >> 17785220

Transvenous cardiac pacing in children: problems and complications during follow-up.

Alpay Celiker1, Osman Başpinar, Tevfik Karagöz.   

Abstract

OBJECTIVE: Transvenous permanent cardiac pacing (TPCP) has become a frequently used therapeutic modality in children. The purpose of this study was to evaluate the outcome of pediatric TPCP regarding problems and complications.
METHODS: Records of 155 patients (mean age 9.2+/-4.7 years) who underwent implantation of TPCP between 1993 and 2003 were reviewed retrospectively. Indications for pacing included atrioventricular block in 76% and sinus node dysfunction in 22% patients. In 92 patients, bradyarrhythmia was secondary to cardiac surgery. Percutaneous subclavian puncture was used for lead implantation in 96% of patients. Pacemakers were placed to the right side of the chest in 84% and in the subpectoral area in 68%. Pacing modes were VVIR in 72%, VDD in 13%, AAIR in 8%, and DDD in 7% of patients at the initial implantation time. Of all electrodes, 95% had steroid elution and 53% had an active fixation mechanism. Mean follow-up period was 37+/-28 (1-120) months.
RESULTS: Forty-five (29%) patients had 21 minor and 45 major complications. Forty-four of 76 revisions were due to lead problems and battery extraction. Most of the lead problems were dislodgment and stretching (n=14). Kaplan Meier analysis of lead survival did not show any difference between lead types. During the follow-up, there were three sudden unexpected deaths.
CONCLUSIONS: In children, TPCP can be used safely and effectively. Although, complications are possible and sometimes lead or generator revision may be necessary, long-term outcome is favorable.

Entities:  

Mesh:

Year:  2007        PMID: 17785220

Source DB:  PubMed          Journal:  Anadolu Kardiyol Derg        ISSN: 1302-8723


  3 in total

1.  Permanent cardiac pacing in pediatric patients.

Authors:  Wael Lotfy; Ranya Hegazy; Osama AbdElAziz; Rodina Sobhy; Hossam Hasanein; Fawzan Shaltout
Journal:  Pediatr Cardiol       Date:  2012-08-12       Impact factor: 1.655

2.  A Case of Transvenous Pacemaker Implantation in a 10-year-old Patient.

Authors:  Jiajia Liu; Yasuyuki Shimada
Journal:  J Rural Med       Date:  2013-12-18

3.  An approach to overcoming specific difficulties with cardiac resynchronization therapy in children.

Authors:  Megan M Constans; Samuel J Asirvatham
Journal:  Indian Pacing Electrophysiol J       Date:  2008-05-01
  3 in total