| Literature DB >> 17764332 |
D Haskel1, Y C Tseng, J C Lang, S Sinogeikin.
Abstract
An instrument has been developed for x-ray magnetic circular dichroism (XMCD) measurements at high pressures and low temperatures. This instrument couples a nonmagnetic copper-beryllium diamond anvil cell featuring perforated diamonds with a helium flow cryostat and an electromagnet. The applied pressure can be controlled in situ using a gas membrane and calibrated using Cu K-edge x-ray absorption fine structure measurements. The performance of this instrument was tested by measuring the XMCD spectra of the Gd(5)Si(2)Ge(2) giant magnetocaloric material.Entities:
Mesh:
Year: 2007 PMID: 17764332 DOI: 10.1063/1.2773800
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523