Soft-copy reading in digital mammography of microcalcifications: diagnostic performance of a 5-megapixel cathode ray tube monitor versus a 3-megapixel liquid crystal display monitor in a clinical setting.

T Uematsu1, M Kasami, Y Uchida.   

Abstract

BACKGROUND: A recent study using dedicated contrast-detail phantoms showed that the image quality of a 3-megapixel (M) monitor can approach that of a 5M monitor in digital mammography.
PURPOSE: To compare a 5M cathode ray tube (CRT) monitor with a 3M liquid crystal display (LCD) monitor for soft-copy reading of digital mammography of microcalcifications in a clinical setting.
MATERIAL AND METHODS: 100 screen-detected microcalcification lesions (34 malignant and 66 benign) without mass that had been evaluated with 11-gauge stereotactic vacuum-assisted breast biopsy or definitive surgery were recruited into the study. One radiologist analyzed the soft-copy mammograms on a 5M CRT monitor and a 3M LCD monitor with 5 months between interpretations and scored the likelihood of malignancy and calcification distribution on a five-point scale. Calcification morphology and breast density were scored on a four-point scale. Positive predictive value (PPV) and negative predictive value (NPV) were calculated on the basis of a Breast Imaging Reporting and Data System. The interpretation time was also measured.
RESULTS: There was no significant difference in the likelihood of malignancy (P = 0.655), calcification morphology (P = 0.168), calcification distribution (P = 0.11), and breast density (P = 0.0608). The PPV and NPV of soft-copy reading on the 5M CRT monitor was 57% (30/53) and 91% (43/47), respectively, identical to the results using the 3M LCD monitor. The total interpretation time averaged 88 s for the 5M CRT monitor and 67 s for the 3M LCD monitor (P<0.0001).
CONCLUSION: Soft-copy reading of a digital mammography of microcalcifications with a 3M LCD monitor was similar in diagnostic performance to a 5M CRT monitor in this study.

Entities:  

Mesh:

Year:  2007        PMID: 17729000     DOI: 10.1080/02841850701370691

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  1 in total

1.  Diagnostic performance of detecting breast cancer on computed radiographic (CR) mammograms: comparison of hard copy film, 3-megapixel liquid-crystal-display (LCD) monitor and 5-megapixel LCD monitor.

Authors:  Takayuki Yamada; Akihiko Suzuki; Nachiko Uchiyama; Noriaki Ohuchi; Shoki Takahashi
Journal:  Eur Radiol       Date:  2008-05-20       Impact factor: 5.315

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.