Literature DB >> 17726602

Comparative performance study of different sample introduction techniques for rapid and precise selenium isotope ratio determination using multi-collector inductively coupled plasma mass spectrometry (MC-ICP/MS).

Nagmeddin Elwaer1, Holger Hintelmann.   

Abstract

The analytical performance of five sample introduction systems, a cross flow nebulizer spray chamber, two different solvent desolvation systems, a multi-mode sample introduction system (MSIS), and a hydride generation (LI2) system were compared for the determination of Se isotope ratio measurements using multi-collector inductively coupled plasma mass spectrometry (MC-ICP/MS). The optimal operating parameters for obtaining the highest Se signal-to-noise (S/N) ratios and isotope ratio precision for each sample introduction were determined. The hydride generation (LI2) system was identified as the most suitable sample introduction method yielding maximum sensitivity and precision for Se isotope ratio measurement. It provided five times higher S/N ratios for all Se isotopes compared to the MSIS, 20 times the S/N ratios of both desolvation units, and 100 times the S/N ratios produced by the conventional spray chamber sample introduction method. The internal precision achieved for the (78)Se/(82)Se ratio at 100 ng mL(-1) Se with the spray chamber, two desolvation, MSIS, and the LI2 systems coupled to MC-ICP/MS was 150, 125, 114, 13, and 7 ppm, respectively. Instrument mass bias factors (K) were calculated using an exponential law correction function. Among the five studied sample introduction systems the LI2 showed the lowest mass bias of -0.0265 and the desolvation system showed the largest bias with -0.0321.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17726602     DOI: 10.1007/s00216-007-1537-z

Source DB:  PubMed          Journal:  Anal Bioanal Chem        ISSN: 1618-2642            Impact factor:   4.142


  1 in total

1.  Novel Approach for the Accurate Determination of Se Isotope Ratio by Multicollector ICP-MS.

Authors:  Jakub Karasiński; Andriy Tupys; Lu Yang; Zoltan Mester; Ludwik Halicz; Ewa Bulska
Journal:  Anal Chem       Date:  2020-11-23       Impact factor: 6.986

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.