Literature DB >> 17724404

Sporicidal activity of disinfectants as one possible cause for bacteria in patient-ready endoscopes.

Norman Miner1, Valerie Harris, Towanda Ebron, Thuy-Dung Cao.   

Abstract

Four endoscopes were cleaned by an experienced endoscopy technician using an enzyme detergent solution with brushing, rinsing with tap water, and then high-level disinfection in an automatic endoscope reprocessing machine using CIDEX orthophthalaldehyde solution (CIDEX OPA). After disinfection, the channels of these patient-ready endoscopes were flushed with sterile neutralizing medium, brushed with a sterile brush, and then flushed again with sterile medium. The effluent from each flush was collected in sterile bottles, immediately returned on ice to a laboratory, and tested for the presence of bacteria. An average of about 200 colony-forming units of bacteria were recovered from each endoscope. Upon staining and microscopic examination, 3 of these colonies were spore-forming bacteria, and 7 colonies were nonspore-forming bacteria. These results suggest that the endoscopes might have been contaminated with a biofilm. Bacterial biofilms have been speculated to commonly occur in endoscopes as a result of the many possible inadequacies of cleaning, disinfecting, rinsing, drying, storage, and other functions associated with the difficulties of reprocessing endoscopes. As one possible cause for a biofilm, three high-level disinfectants (CIDEX activated dialdehyde solution, CIDEX OPA, and Aldahol high-level disinfectant) were tested for their sporicidal activity against high-protein or low-protein cultures of spore-forming bacteria in suspension. The potential importance of killing spore-forming bacteria within a practical exposure time in order to prevent the formation of biofilms is discussed.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17724404     DOI: 10.1097/01.SGA.0000287201.98483.43

Source DB:  PubMed          Journal:  Gastroenterol Nurs        ISSN: 1042-895X            Impact factor:   0.978


  2 in total

Review 1.  Is peracetic acid suitable for the cleaning step of reprocessing flexible endoscopes?

Authors:  Günter Kampf; Patricia M Fliss; Heike Martiny
Journal:  World J Gastrointest Endosc       Date:  2014-09-16

2.  Effect of Different Disinfectants on Bacterial Aerosol Diversity in Poultry Houses.

Authors:  Linlin Jiang; Meng Li; Jinxiu Tang; Xiaoyu Zhao; Jianlong Zhang; Hongwei Zhu; Xin Yu; Youzhi Li; Tao Feng; Xingxiao Zhang
Journal:  Front Microbiol       Date:  2018-09-11       Impact factor: 5.640

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.