Deadspace estimation from CO2 versus molar mass measurements in infants.

C Thamrin1, P Latzin, L Sauteur, T Riedel, G L Hall, U Frey.   

Abstract

BACKGROUND: Estimation of respiratory deadspace is often based on the CO2 expirogram, however presence of the CO2 sensor increases equipment deadspace, which in turn influences breathing pattern and calculation of lung volume. In addition, it is necessary to correct for the delay between the sensor and flow signals. We propose a new method for estimation of effective deadspace using the molar mass (MM) signal from an ultrasonic flowmeter device, which does not require delay correction. We hypothesize that this estimation is correlated with that calculated from the CO2 signal using the Fowler method.
METHODS: Breath-by-breath CO2, MM and flow measurements were made in a group of 77 term-born healthy infants. Fowler deadspace (Vd,Fowler) was calculated after correcting for the flow-dependent delay in the CO2 signal. Deadspace estimated from the MM signal (Vd,MM) was defined as the volume passing through the flowhead between start of expiration and the 10% rise point in MM.
RESULTS: Correlation (r = 0.456, P < 0.0001) was found between Vd,MM and Vd,Fowler averaged over all measurements, with a mean difference of -1.4% (95% CI -4.1 to 1.3%). Vd,MM ranged from 6.6 to 11.4 ml between subjects, while Vd,Fowler ranged from 5.9 to 12.0 ml. Mean intra-measurement CV over 5-10 breaths was 7.8 +/- 5.6% for Vd,MM and 7.8 +/- 3.7% for Vd,Fowler. Mean intra-subject CV was 6.0 +/- 4.5% for Vd,MM and 8.3 +/- 5.9% for Vd,Fowler. Correcting for the CO2 signal delay resulted in a 12% difference (P = 0.022) in Vd,Fowler. Vd,MM could be obtained more frequently than Vd,Fowler in infants with CLD, with a high variability.
CONCLUSIONS: Use of the MM signal provides a feasible estimate of Fowler deadspace without introducing additional equipment deadspace. The simple calculation without need for delay correction makes individual adjustment for deadspace in FRC measurements possible. This is especially important given the relative large range of deadspace seen in this homogeneous group of infants. (c) 2007 Wiley-Liss, Inc.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17722053     DOI: 10.1002/ppul.20683

Source DB:  PubMed          Journal:  Pediatr Pulmonol        ISSN: 1099-0496


  4 in total

1.  Tidal volume single breath washout of two tracer gases--a practical and promising lung function test.

Authors:  Florian Singer; Georgette Stern; Cindy Thamrin; Oliver Fuchs; Thomas Riedel; Per Gustafsson; Urs Frey; Philipp Latzin
Journal:  PLoS One       Date:  2011-03-10       Impact factor: 3.240

2.  Impact of Software Settings on Multiple-Breath Washout Outcomes.

Authors:  Selina Summermatter; Florian Singer; Philipp Latzin; Sophie Yammine
Journal:  PLoS One       Date:  2015-07-13       Impact factor: 3.240

3.  Cohort profile: the Bern infant lung development cohort.

Authors:  Oliver Fuchs; Philipp Latzin; Claudia E Kuehni; Urs Frey
Journal:  Int J Epidemiol       Date:  2011-01-13       Impact factor: 7.196

Review 4.  Current methodological and technical limitations of time and volumetric capnography in newborns.

Authors:  Gerd Schmalisch
Journal:  Biomed Eng Online       Date:  2016-08-30       Impact factor: 2.819

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.