Literature DB >> 17721504

Retinal nerve fibre layer imaging compared with histological measurements in a human eye.

E Z Blumenthal1, R S Parikh, J Pe'er, M Naik, E Kaliner, M J Cohen, S Prabakaran, M Kogan, R Thomas.   

Abstract

PURPOSE: A feasibility study comparing retinal nerve fibre layer (RNFL) thickness values obtained with imaging devices against RNFL thickness measurements obtained histologically in a human eye.
DESIGN: A single patient scheduled for orbital exenteration, who still possessed a healthy functioning eye.
METHODS: Before surgery, the eye was imaged using optical coherence tomography (OCT) and scanning laser polarimetry (SLP). After orbital exenteration, the globe was sectioned, and 100 equidistant RNFL thickness measurements were obtained for each of four concentric rings centred on the optic disc, with diameters of 3.0, 3.5, 4.0, and 4.5 mm.
RESULTS: RNFL thickness was found to be inversely related to the distance from the centre of the optic disc along each radial meridian. Peripapillary RNFL thickness was found comparable for histology, OCT, and SLP. RNFL thickness measured histologically confirmed a 'double hump' pattern, peaking at the superior and inferior poles. Histologically derived RNFL thickness (microm), at 3.0 and 3.5 mm diameter ring ranged between 30-135 and 25-115 respectively. In comparison, the 3.0 mm diameter GDx data ranged between 25 and 100, and the 3.4 mm diameter OCT data between 40 and 175.
CONCLUSIONS: Imaging data appear qualitatively similar when compared to the histologically derived data. Quantitative differences may be partly due to scaling differences and histological artefacts. The histological analysis approach demonstrated in this study can potentially serve to validate imaging-derived data, as well as help improve our understanding of RNFL loss in glaucoma.

Entities:  

Mesh:

Year:  2007        PMID: 17721504     DOI: 10.1038/sj.eye.6702942

Source DB:  PubMed          Journal:  Eye (Lond)        ISSN: 0950-222X            Impact factor:   3.775


  20 in total

1.  From Imaging the Brain to Imaging the Retina: Optical Coherence Tomography (OCT) in Schizophrenia.

Authors:  Carlos Schönfeldt-Lecuona; Thomas Kregel; Arno Schmidt; Elmar H Pinkhardt; Florian Lauda; Jan Kassubek; Bernhard J Connemann; Roland W Freudenmann; Maximilian Gahr
Journal:  Schizophr Bull       Date:  2015-06-05       Impact factor: 9.306

2.  Transitioning from stratus OCT to cirrus OCT: a comparison and a proposed equation to convert central subfield macular thickness measurements in healthy subjects.

Authors:  Gelareh Abedi; Payal Patal; Gheorghe Doros; Manju L Subramanian
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2011-07-01       Impact factor: 3.117

3.  Advances in the Structural Evaluation of Glaucoma with Optical Coherence Tomography.

Authors:  Daniel Meira-Freitas; Renato Lisboa; Felipe A Medeiros
Journal:  Curr Ophthalmol Rep       Date:  2013-06-01

4.  Retinal Nerve Fibre Layer and Macular Thinning in Spinocerebellar Ataxia and Cerebellar Multisystem Atrophy.

Authors:  John H Pula; Vernon L Towle; Victoria M Staszak; Dingcai Cao; Jacqueline T Bernard; Christopher M Gomez
Journal:  Neuroophthalmology       Date:  2011-06

5.  Glaucoma classification model based on GDx VCC measured parameters by decision tree.

Authors:  Mei-Ling Huang; Hsin-Yi Chen
Journal:  J Med Syst       Date:  2009-07-04       Impact factor: 4.460

6.  Evaluation of retinal nerve fiber layer thickness in patients with retinitis pigmentosa using spectral-domain optical coherence tomography.

Authors:  Anastasios Anastasakis; Mohamed A Genead; J Jason McAnany; Gerald A Fishman
Journal:  Retina       Date:  2012-02       Impact factor: 4.256

7.  Spectral domain optical coherence tomography cross-sectional image of optic nerve head during intraocular pressure elevation.

Authors:  Ji Young Lee; You Kyung Lee; Jung Il Moon; Myoung Hee Park
Journal:  Int J Ophthalmol       Date:  2014-12-18       Impact factor: 1.779

8.  Retinal nerve fiber layer thickness in subgroups of multiple sclerosis, measured by optical coherence tomography and scanning laser polarimetry.

Authors:  Theodora A M Siepman; Marijke Wefers Bettink-Remeijer; Rogier Q Hintzen
Journal:  J Neurol       Date:  2010-05-12       Impact factor: 4.849

9.  Comparison of retinal nerve fiber layer measurements using time domain and spectral domain optical coherent tomography.

Authors:  O'Rese J Knight; Robert T Chang; William J Feuer; Donald L Budenz
Journal:  Ophthalmology       Date:  2009-04-22       Impact factor: 12.079

10.  A comparison of optic nerve head morphology viewed by spectral domain optical coherence tomography and by serial histology.

Authors:  Nicholas G Strouthidis; Jonathan Grimm; Galen A Williams; Grant A Cull; David J Wilson; Claude F Burgoyne
Journal:  Invest Ophthalmol Vis Sci       Date:  2009-10-29       Impact factor: 4.799