Literature DB >> 17721025

Efficacy of coronary venoplasty for left ventricular lead implantation.

Yoshimitsu Soga1, Kenji Ando, Takashi Yamada, Masahiko Goya, Shinichi Shirai, Koyu Sakai, Masashi Iwabuchi, Hitoshi Yasumoto, Hiroyoshi Yokoi, Hideyuki Nosaka, Masakiyo Nobuyoshi.   

Abstract

BACKGROUND: Optimal target vein selection for the pacing lead results in a better outcome, but left ventricular (LV) variability limits this selection. The aim of the present study was to investigate the efficacy of coronary venoplasty for insertion of a LV lead. METHODS AND
RESULTS: Transvenous LV lead placement was attempted in 208 consecutive patients from November 2002 to January 2007, with success in 206 patients (99%). Retrospective analysis of the cardiac resynchronization therapy system implantation showed that 4 of the 206 patients (1.9%) required coronary venoplasty for insertion of the pacing lead implant. Using coronary balloon angioplasty catheters of 2.5 mm (2 patients), 3.0 mm (1 patient) and 4.0 mm (1 patient), each target vein was dilated. Of 4 patients, stenoses in 3 were dilated by balloon angioplasty only. However, focal stenosis of 1 patient was not able to be dilated due to severe stenosis. Therefore, part of the lesion was sharpened by rotational atherectomy and the stenosis was successfully dilated. The LV pacing lead could then be inserted and no complications occurred.
CONCLUSION: Venoplasty for stenosis was effective in allowing deployment of a LV lead into a target vein in some patients. The safety and complications of the procedure remain unclear.

Entities:  

Mesh:

Year:  2007        PMID: 17721025     DOI: 10.1253/circj.71.1442

Source DB:  PubMed          Journal:  Circ J        ISSN: 1346-9843            Impact factor:   2.993


  1 in total

1.  Implantation of a cardiac resynchronization therapy device using the anchor balloon technique in a patient with a tortuous coronary sinus branch.

Authors:  Yu Kumagai; Takanori Arimoto; So Yamauchi; Daisuke Kutsuzawa; Hayato Tsuchiya; Masafumi Watanabe
Journal:  HeartRhythm Case Rep       Date:  2018-04-30
  1 in total