Literature DB >> 17720081

Intraocular lenses with surface aspherization: Interferometric study.

Rita Mencucci1, Ugo Menchini, Roberto Volpe, Maurizio Vannoni, Giuseppe Molesini.   

Abstract

PURPOSE: To accurately evaluate in an optical laboratory 2 different types of intraocular lenses (IOLs).
SETTING: Department of Oto-Neuro-Ophthalmological Surgical Sciences and CNR-Institute for Applied Optics, Florence Italy.
METHODS: In this study, optical testing facilities based on interferometry were used. The IOLs were inspected with laser light, and interference patterns were acquired and processed, obtaining the physical parameters accounting for the lens behavior. Comparison of the respective performances within the pseudophakic eye was provided in terms of modulation transfer function (MTF) using lens diameters from 3.0 mm to 5.0 mm. The data were inserted in a mathematical model of the eye, and the relevant merit functions were computed with a ray-tracing program.
RESULTS: Sample sets of both all-spherical IOLs and IOLs with aspherical surfaces were inspected. The interferometric maps showed clear evidence of the opposite wave aberration introduced by the different IOL types. With a 3.0 mm pupil diameter, both IOL types performed as almost ideal lenses, while with a 5.0 mm diameter, IOLs with aspherical surfaces effectively compensated for the spherical aberration of the cornea and had a significantly higher MTF.
CONCLUSIONS: The interferometric approach to IOL testing provided accurate data on the physical parameters of the IOL. With a measuring and computing chain that includes eye modeling and ray tracing, it was possible to analyze in detail the process by which IOLs with aspherical surfaces compensate for the spherical aberration of the cornea under photopic, mesopic, and scotopic conditions.

Entities:  

Mesh:

Year:  2007        PMID: 17720081     DOI: 10.1016/j.jcrs.2007.05.028

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  1 in total

1.  Individual IOL surface topography analysis by the WaveMaster Reflex UV.

Authors:  Marc Kannengießer; Achim Langenbucher; Edgar Janunts
Journal:  Biomed Res Int       Date:  2013-05-28       Impact factor: 3.411

  1 in total