Motion artifacts of extended high frame rate imaging.

Jing Wang1, Jian-yu Lu.   

Abstract

Based on the high frame rate (HFR) imaging method developed in our lab, an extended high frame rate imaging method with various transmission schemes was developed recently. In this method, multiple, limited-diffraction array beams or steered plane wave transmissions are used to increase image resolution and field of view as well as to reduce sidelobes. Furthermore, the multiple, limited-diffraction array beam transmissions can be approximated with square-wave aperture weightings, allowing one or two transmitters to be used with a multielement array transducer to simplify imaging systems. By varying the number of transmissions, the extended HFR imaging method allows a continuous trade-off between image quality and frame rate. Because multiple transmissions are needed to obtain one frame of image for the method, motion could cause phase misalignment and thus produce artifacts, reducing image contrast and resolution and leading to an inaccurate clinical interpretation of images. Therefore, it is important to study how motion affects the method and provide a useful guidance of using the method properly in various applications. In this paper, computer simulations, in vitro and in vivo experiments were performed to study the effects of motion on the method in different conditions. Results show that a number of factors may affect the motion effects. However, it was found that the extended HFR imaging method is not sensitive to the motions commonly encountered in the clinical applications, as is demonstrated by an in vivo heart experiment, unless the number of transmissions is large and objects are moving at a high velocity near the surface of a transducer.

Mesh:

Year:  2007        PMID: 17718320     DOI: 10.1109/tuffc.2007.391

Source DB:  PubMed          Journal:  IEEE Trans Ultrason Ferroelectr Freq Control        ISSN: 0885-3010            Impact factor:   2.725


  3 in total

1.  Pulse wave imaging using coherent compounding in a phantom and in vivo.

Authors:  Iason Zacharias Apostolakis; Matthew D J McGarry; Ethan A Bunting; Elisa E Konofagou
Journal:  Phys Med Biol       Date:  2016-12-21       Impact factor: 3.609

2.  High-Frame-Rate Doppler Ultrasound Using a Repeated Transmit Sequence.

Authors:  Anthony S Podkowa; Michael L Oelze; Jeffrey A Ketterling
Journal:  Appl Sci (Basel)       Date:  2018-02-01       Impact factor: 2.679

3.  High-contrast ultrafast imaging of the heart.

Authors:  Clement Papadacci; Mathieu Pernot; Mathieu Couade; Mathias Fink; Mickael Tanter
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2014-02       Impact factor: 2.725

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.