Introduction of infection control measures to reduce infection associated with implantable pain therapy devices.

Abram H Burgher1, Crawford F Barnett, Jon B Obray, W David Mauck.   

Abstract

INTRODUCTION: Implantable pain therapy devices for chronic pain include spinal cord stimulators (SCS) and intrathecal drug delivery systems (IDDS). A number of different complications can occur after implantation of these devices, but among the most serious is infection. Based on Centers for Disease Control and Prevention guidelines for prevention of surgical site infection, published literature on infection risk with implantable pain therapy devices, and recommendations from groups within our own our institution, we introduced infection control measures for all patients receiving either SCS or IDDS.
METHODS: After approval from the Institutional Review Board, we performed a retrospective review of patients undergoing primary implantation of SCS or IDDS before and after introduction at our institution of safety measures designed to reduce device-related infection. We compared infection incidence and compliance to infection precautions before and after introduction of these measures.
RESULTS: Thirty-four SCS or IDDS were implanted before implementation of the infection control measures and 58 were placed after. Five device-related infections occurred. Adherence to most infection precautions improved during the study period, but 100% compliance was seen only with venue used for implantation. Infection incidence declined after introduction of the safety measures, but the reduction was not statistically significant.
CONCLUSIONS: Introduction of infection control measures for implantable pain therapy devices improved adherence to most infection precautions in our practice. Lack of specific documentation could have hindered practice surveillance within our group. A tool to document performance of infection control measures would be useful not only as a marker of compliance but could also serve as a reminder to perform certain safety measures.

Entities:  

Mesh:

Year:  2007        PMID: 17714108     DOI: 10.1111/j.1533-2500.2007.00142.x

Source DB:  PubMed          Journal:  Pain Pract        ISSN: 1530-7085            Impact factor:   3.183


  4 in total

Review 1.  Peripheral nerve stimulation for the treatment of primary headache.

Authors:  Pyungbok Lee; Billy K Huh
Journal:  Curr Pain Headache Rep       Date:  2013-03

2.  Adverse Events and Complications Associated With Intrathecal Drug Delivery Systems: Insights From the Manufacturer and User Facility Device Experience (MAUDE) Database.

Authors:  Vasudha Goel; Yan Yang; Siddak Kanwar; Ratan K Banik; Amol M Patwardhan; Mohab Ibrahim; Eellan Sivanesan; Hariharan Shankar
Journal:  Neuromodulation       Date:  2020-12-11

3.  Systematic Review and Meta-Analysis of Preoperative Antisepsis with Combination Chlorhexidine and Povidone-Iodine.

Authors:  Benjamin M Davies; Hiren C Patel
Journal:  Surg J (N Y)       Date:  2016-08-10

4.  Postoperative Infections Associated With Prolonged Spinal Cord Stimulation Trial Duration (PROMISE RCT).

Authors:  Richard North; Mehul J Desai; Johan Vangeneugden; Christian Raftopoulos; Tony Van Havenbergh; Marc Deruytter; Jean-Michel Remacle; Jane Shipley; Ye Tan; Mary Jo Johnson; Carine Van den Abeele; Philippe Rigoard
Journal:  Neuromodulation       Date:  2020-04-08
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.