Literature DB >> 17712199

Coronary stenting with M-Guard: feasibility and safety porcine trial.

Edo Kaluski1, Adam Groothuis, Marc Klapholz, Philip Seifart, Elazar Edelman.   

Abstract

UNLABELLED: M-Guard is an ultra-thin polymer mesh sleeve attached to the external stent surface. It is designed to minimize distal embolization during coronary, renal, carotid and vein graft stenting. The polymer net could also serve as a platform for more uniform drug delivery. AIM: To evaluate coronary M-Guard stent deliverability and safety (stent thrombosis and restenosis) in comparison to bare-metal stents (BMS) in a porcine model of percutaneous coronary interventions (PCIs).
METHODS: Under general anesthesia using percutaneous technique, 6 swine received a total of 18 coronary stents: 5 BMS and 13 M-Guard-BMS. Quantitative coronary angiography (QCA) was obtained immediately prior to and post-PCI, and at 30 days post-stenting. At 30 days, all animals were sacrificed and hearts were sent to a core lab for coronary histology and histomorphometry. Primary endpoints were 30-day QCA percent diameter stenosis, late luminal loss and minimal luminal diameter (MLD). Secondary endpoints were procedural success, 30-day mortality and stent thrombosis. Exploratory endpoints were histology and histomorphometric analysis performed at 30 days on M-Guard stented segments.
RESULTS: All stents were delivered successfully. There were no procedural complications or porcine morbidity or mortality at 30 days. The M-Guard and BMS displayed similar results of MLD, late luminal loss and percent diameter stenosis at 1 month.
CONCLUSION: When compared to BMS, PCI with M-Guard-BMS is feasible, safe and yields similar inflammatory and restenotic response.

Entities:  

Mesh:

Year:  2007        PMID: 17712199

Source DB:  PubMed          Journal:  J Invasive Cardiol        ISSN: 1042-3931            Impact factor:   2.022


  2 in total

1.  A new device to seal large coronary aneurysms: a case report.

Authors:  Gian Battista Danzi; Guido Angelo Pomidossi; Filippo Casolo; Marco Centola; Roberto Ferraresi; Chaim Lotan
Journal:  J Med Case Rep       Date:  2010-08-03

2.  Ultrasound-guided femoral approach for coronary angiography and interventions in the porcine model.

Authors:  Grigorios Tsigkas; Georgios Vasilagkos; Alexandros Tousis; Michail Theofanis; Anastasios Apostolos; Ioannis Spyridonidis; Leonidas Goudas; Georgios Karpetas; Athanasios Moulias; Christos S Katsouras; Panagiotis Kitrou; Virginia Mplani; Anargyros N Moulas; Dimitrios Karnabatidis; Periklis Davlouros
Journal:  Sci Rep       Date:  2022-08-17       Impact factor: 4.996