Cement leakage into the posterior spinal canal during balloon kyphoplasty: a case report.

S Becker1, J Meissner, A Tuschel, A Chavanne, M Ogon.   

Abstract

We report a case of cement leakage into the posterior spinal canal due to inadvertent pedicle perforation during balloon kyphoplasty. The leakage was corrected immediately without any sequelae. Features seen on radiography and the minimally invasive procedure used for removal are described. The postoperative radiographs of 100 consecutive patients treated with balloon kyphoplasty were subsequently reviewed. Only one patient had a similar leakage but had no neurological complications.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17709866     DOI: 10.1177/230949900701500220

Source DB:  PubMed          Journal:  J Orthop Surg (Hong Kong)        ISSN: 1022-5536            Impact factor:   1.118


  7 in total

1.  [Vena cava cement embolism after percutaneous kyphoplasty. Interventional radiological recovery in metastasized urothelial cancer].

Authors:  R Schmid; K Efinger; C Schulz; C Sparwasser; A Martinschek
Journal:  Urologe A       Date:  2014-01       Impact factor: 0.639

2.  Treatment of severe vertebral body compression fractures with percutaneous vertebroplasty.

Authors:  Casey Young; Peter L Munk; Manraj K Heran; Michael D Lane; Huy B Q Le; Steven Lee; M Badii; Paul W Clarkson; Hugue A Ouellette; Ouellette C Hugue
Journal:  Skeletal Radiol       Date:  2011-03-27       Impact factor: 2.199

3.  Comparison of the results of balloon kyphoplasty performed at different times after injury.

Authors:  Gun Soek Oh; Hyeun Sung Kim; Chang Il Ju; Seok Won Kim; Seung Myung Lee; Ho Shin
Journal:  J Korean Neurosurg Soc       Date:  2010-03-31

4.  Epidural cement leakage through pedicle violation after balloon kyphoplasty causing paraparesis in osteoporotic vertebral compression fractures - a report of two cases.

Authors:  Si-Young Park; Hitesh N Modi; Seung-Woo Suh; Jae-Young Hong; Won Noh; Jae-Hyuk Yang
Journal:  J Orthop Surg Res       Date:  2010-08-06       Impact factor: 2.359

5.  Applications of memory alloy stent in vertebral fractures.

Authors:  Yang Yimin; Zhang Zhi; Ren ZhiWei; Ma Wei; Rajiv Kumar Jha
Journal:  Med Sci Monit Basic Res       Date:  2014-05-28

6.  Surgical Removal of Circumferentially Leaked Polymethyl Methacrylate in the Epidural Space of the Thoracic Spine after Percutaneous Vertebroplasty.

Authors:  Kenichiro Kita; Yoichiro Takata; Kosaku Higashino; Kazuta Yamashita; Fumitake Tezuka; Toshinori Sakai; Akihiro Nagamachi; Koichi Sairyo
Journal:  Surg J (N Y)       Date:  2017-02-10

7.  Complications and safety aspects of kyphoplasty for osteoporotic vertebral fractures: a prospective follow-up study in 102 consecutive patients.

Authors:  Yohan Robinson; Sven Kevin Tschöke; Philip F Stahel; Ralph Kayser; Christoph E Heyde
Journal:  Patient Saf Surg       Date:  2008-01-15
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.