Transcutaneous bilirubin measurement: evaluation of Bilitest.

Giovanna Bertini1, Simone Pratesi, Elena Cosenza, Carlo Dani.   

Abstract

OBJECTIVE: The early discharge of neonates from hospitals makes transcutaneous measurement of total serum bilirubin concentration a useful tool to monitor neonatal jaundice. The objective of this study was to evaluate the Bilitest BB 77trade mark (Bertocchi SRL Elettromedicali, Cremona, Italy), a new device for noninvasive transcutaneous total bilirubin measurement.
METHODS: We studied 241 newborn infants > or =32 weeks of gestation admitted to the Neonatal Nursery of Careggi University Hospital, Florence. These infants had total serum bilirubin (TSB) levels measured by a standard laboratory test [corrected] as part of their normal care, and transcutaneous bilirubin (TcB) levels were obtained within 10 min after heel pricking.
RESULTS: There was a good correlation between TSB and TcB values. The linear regression plot showed a general underestimation of Bilitest BB 77 measurement compared to standard laboratory test; the negative difference between Bilitest BB 77 values and TSB increased at higher bilirubin levels, as confirmed by the Bland-Altman error plot. To visualize the accuracy of the Bilitest BB 77 measurements, ROC curves plot sensitivity versus specificity and the maximum range of difference between Bilitest BB 77 and TSB were considered. Pearson correlation analysis demonstrated that Bilitest BB 77 accuracy was independent of gestational and postnatal age of newborns.
CONCLUSION: We conclude that the Bilitest BB 77, a new device for bilirubin transcutaneous measurement, shows a good correlation with standard laboratory method and has the merit to be quite inexpensive and does not need calibration with consumable or disposable parts. However, because TcB measurements with Bilitest BB 77 underestimated STB levels particularly at STB > or =12 mg/dl, serum bilirubin measurements are still required when treatment with phototherapy or exchange transfusion is being considered. (c) 2007 S. Karger AG, Basel

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17703104     DOI: 10.1159/000107351

Source DB:  PubMed          Journal:  Neonatology        ISSN: 1661-7800            Impact factor:   4.035


  3 in total

1.  Photoacoustic microscopy of bilirubin in tissue phantoms.

Authors:  Yong Zhou; Chi Zhang; Da-Kang Yao; Lihong V Wang
Journal:  J Biomed Opt       Date:  2012-12       Impact factor: 3.170

2.  Measuring transcutaneous bilirubin: a comparative analysis of three devices on a multiracial population.

Authors:  Francesco Raimondi; Silvia Lama; Francesca Landolfo; Maria Sellitto; Angela Carla Borrelli; Rosalba Maffucci; Paola Milite; Letizia Capasso
Journal:  BMC Pediatr       Date:  2012-06-14       Impact factor: 2.125

3.  Evaluation of transcutaneous bilirubinometer (DRAEGER JM 103) use in Zimbabwean newborn babies.

Authors:  Gwendoline Lilly Tanyaradzwa Chimhini; Simbarashe Chimhuya; Vasco Chikwasha
Journal:  Matern Health Neonatol Perinatol       Date:  2018-01-18
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.