Literature DB >> 1770107

Elimination of caffeine interference in high-performance liquid chromatographic determination of cotinine in human plasma.

S Pichini1, I Altieri, R Pacifici, M Rosa, P Zuccaro.   

Abstract

A high-performance liquid chromatographic method with ultraviolet photometric detection for the determination of cotinine in human plasma was described. The use of a 30-cm reversed-phase column and of a mobile phase consisting of water-methanol-0.1 M sodium acetate-acetonitrile (72:21:5.6:1.4, v/v), pH 4.1, eliminated caffeine interference. A simplified solid-phase extraction procedure was also performed for plasma samples.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1770107     DOI: 10.1016/0378-4347(91)80363-h

Source DB:  PubMed          Journal:  J Chromatogr


  1 in total

Review 1.  Chemical analyses as validators in smoking cessation programs.

Authors:  D D Gilbert
Journal:  J Behav Med       Date:  1993-06
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.