Literature DB >> 17697818

Frequency of stent fracture as a cause of coronary restenosis after sirolimus-eluting stent implantation.

Sang-Hee Lee1, Jong-Seon Park, Dong-Gu Shin, Young-Jo Kim, Gue-Ru Hong, Woong Kim, Bong-Sup Shim.   

Abstract

Stent fracture (SF) was suggested as a cause of restenosis after sirolimus-eluting stent (SES) implantation. This study was performed to evaluate the incidence and characteristics of SF to determine its contribution to restenosis in patients with in-stent restenosis (ISR) after SES implantation. From May 2003 to February 2006, SESs were used for percutaneous coronary intervention in 868 patients with 1,109 coronary narrowings. Follow-up coronary angiography was performed in 366 patients (42%), and 26 ISR lesions were observed. These patients were enrolled in this study. SF was divided into 3 types as avulsion, collapse, and partial based on the findings of fluoroscopy, coronary angiography, and intravascular ultrasound study. Of 26 patients with ISR lesions, SF was identified in 10. SF types were avulsion (5 patients), collapse (2 patients), and partial (3 patients). SF was identified at the midshaft (7 patients) and overlap sites (3 patients) of stents. SF was not observed in the 30 patients with ISR after bare-metal Bx Velocity stent implantation. Four patients with SF were treated with paclitaxel-eluting stents. In conclusion, SF is 1 of the leading causes of ISR after SES implantation. Careful fluoroscopic examination is necessary at the time of follow-up angiography to identify this problem.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17697818     DOI: 10.1016/j.amjcard.2007.03.073

Source DB:  PubMed          Journal:  Am J Cardiol        ISSN: 0002-9149            Impact factor:   2.778


  18 in total

1.  Evaluation of in-stent neointimal tissue components using integrated backscatter intravascular ultrasound: comparison of drug-eluting stents and bare-metal stents.

Authors:  Yoshitaka Muraoka; Shinjo Sonoda; Kuninobu Kashiyama; Fumihiko Kamezaki; Yuki Tsuda; Masaru Araki; Masahiro Okazaki; Yutaka Otsuji
Journal:  Int J Cardiovasc Imaging       Date:  2011-12-18       Impact factor: 2.357

2.  Prevalence and prognosis of coronary stent gap detected by multi-detector CT: a follow-up study.

Authors:  Xinghua Zhang; Li Yang; Haiyue Ju; Fan Zhang; Jian Wu; Bin He; Yundai Chen
Journal:  Eur Radiol       Date:  2012-04-27       Impact factor: 5.315

Review 3.  Are we aware of stent fracture?

Authors:  R D Acar; M Bulut; M Akcakoyun
Journal:  Herz       Date:  2013-10-25       Impact factor: 1.443

4.  Coronary stent fracture.

Authors:  Benjamin Hibbert; Edward R O'Brien
Journal:  CMAJ       Date:  2011-03-28       Impact factor: 8.262

5.  Fracture of zotarolimus-eluting stent after implantation.

Authors:  Meng-Chieh Wu; Chen-Chuan Cheng; Tsuei-Yuan Huang
Journal:  Tex Heart Inst J       Date:  2009

6.  Strain-induced accelerated asymmetric spatial degradation of polymeric vascular scaffolds.

Authors:  Pei-Jiang Wang; Nicola Ferralis; Claire Conway; Jeffrey C Grossman; Elazer R Edelman
Journal:  Proc Natl Acad Sci U S A       Date:  2018-02-26       Impact factor: 11.205

7.  Effect of working environment and procedural strategies on mechanical performance of bioresorbable vascular scaffolds.

Authors:  Pei-Jiang Wang; Farhad Rikhtegar Nezami; Maysam B Gorji; Francesca Berti; Lorenza Petrini; Tomasz Wierzbicki; Francesco Migliavacca; Elazer R Edelman
Journal:  Acta Biomater       Date:  2018-10-17       Impact factor: 8.947

8.  CT Imaging of Coronary Stents: Past, Present, and Future.

Authors:  Andreas H Mahnken
Journal:  ISRN Cardiol       Date:  2012-09-11

9.  An unusual case of ST-segment elevation myocardial infarction following a late bare-metal stent fracture in a native coronary artery: a case report.

Authors:  Giovanni Minardi; Paolo G Pino; Marco Stefano Nazzaro; Herribert Pavaci; Martina Sordi; Cesare Greco; Carlo Gaudio
Journal:  J Med Case Rep       Date:  2009-11-24

10.  Pathologic Etiologies of Late and Very Late Stent Thrombosis following First-Generation Drug-Eluting Stent Placement.

Authors:  Fumiyuki Otsuka; Masataka Nakano; Elena Ladich; Frank D Kolodgie; Renu Virmani
Journal:  Thrombosis       Date:  2012-11-21
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.