Physical properties, durability, and energy-dissipation function of dual-density orthotic materials used in insoles for diabetic patients.

James W Brodsky1, Fabian E Pollo, Domingo Cheleuitte, Brian S Baum.   

Abstract

BACKGROUND: Patients with neuropathic conditions may develop plantar bony deformities through neuropathic collapse, frequently placing the skin and soft tissues at risk. Orthoses have been used to accommodate and distribute plantar pressures over a large surface area, thereby minimizing peak loading pressures in small regions and reducing the risk of ulceration.
METHODS: A previously described bony prominence model (Brodsky et al.) was used to test the pressure-absorbing and force-transmission properties of various orthotic material combinations used in our outpatient clinic. Six materials were tested in five combinations of materials for their compressive properties: [MS]: medium plastazote (M) + soft plastazote (S); [MN]: medium plastazote (M) + nickelplast (N); [NP] nickelplast (N) + Poron (P); [MO] medium plastazote (M) + Spenco (O); and [MC] medium plastazote (M) + P-cell (C). Materials were tested for 100,000 cycles using a materials-testing system (MTS) apparatus (MTS Systems Corporation, Cary, NC) and software. Stress-strain curves comparing the measured peak pressure to the elastic deformation, or the percentage of compression a material experiences with respect to its original thickness, were plotted for each orthotic combination.
RESULTS: For MS, MN, MO, and to a lesser extent, MC, a trend was noted for decreased elastic deformation with increased testing. Additionally, the peak pressures before and after testing for each 10,000 testing cycle for each of the orthotic combinations were plotted. For both MN and NP, no demonstrable difference was noted in the peak pressures in the pretesting and post-testing for the 100,000 cycles. The MO showed a trend for increased peak pressures after each testing cycle. Both the MC and MS peak pressures markedly increased with respect to pretesting value. Also, the MN, MO, and MS all showed an overall trend for increased load cell values with increasing cycles at fast loading.
CONCLUSIONS: These data showed that some orthotic combinations are more effective than others at reducing peak pressures during compression testing using our bony prominence model. Further studies are needed to test the orthotic combinations for shear and combined shear and compression modes.

Entities:  

Mesh:

Year:  2007        PMID: 17697652     DOI: 10.3113/FAI.2007.0880

Source DB:  PubMed          Journal:  Foot Ankle Int        ISSN: 1071-1007            Impact factor:   2.827


  3 in total

1.  An apparatus to quantify anteroposterior and mediolateral shear reduction in shoe insoles.

Authors:  Barry Belmont; Yancheng Wang; Peethambaran Ammanath; James S Wrobel; Albert Shih
Journal:  J Diabetes Sci Technol       Date:  2013-03-01

2.  A novel workflow to fabricate a patient-specific 3D printed accommodative foot orthosis with personalized latticed metamaterial.

Authors:  Yuri F Hudak; Jing-Sheng Li; Scott Cullum; Brian M Strzelecki; Chris Richburg; G Eli Kaufman; Daniel Abrahamson; Jeffrey T Heckman; Beth Ripley; Scott Telfer; William R Ledoux; Brittney C Muir; Patrick M Aubin
Journal:  Med Eng Phys       Date:  2022-04-14       Impact factor: 2.356

Review 3.  Footwear and insole design features for offloading the diabetic at risk foot-A systematic review and meta-analyses.

Authors:  Richard Collings; Jennifer Freeman; Jos M Latour; Joanne Paton
Journal:  Endocrinol Diabetes Metab       Date:  2020-04-11
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.