Literature DB >> 17696626

Fenestrated and branched grafts for para-anastomotic aortic aneurysm repair.

Peter Ziegler1, Theodossios P Perdikides, Efthimios D Avgerinos, Thomas Umscheid, Wolf J Stelter.   

Abstract

PURPOSE: To investigate the use of fenestration and branch artery stenting during endovascular stent-graft repair of para-anastomotic aneurysms (PAA).
METHODS: A retrospective review was conducted of 9 patients (all men; mean age 71 years, range 60-80) who received custom-designed fenestrated endoprostheses for PAA repair. Eight tubular fenestrated devices and 1 composite device (fenestrated tube plus modular bifurcated body) with a total of 31 fenestrations were used.
RESULTS: The mean operating time was 318+/-93 minutes (range 220-485); the mean fluoroscopy time was 77+/-38 minutes (range 39-158), during which a mean 121+/-81 mL (range 33-300) of contrast was used. Technical success was achieved in all cases. Over a mean follow-up of 12+/-5.5 months (range 6-24), 1 secondary intervention was carried out due to a break in a side branch stent-graft; 2 transient renal impairments and 1 permanent renal insufficiency unrelated to renal artery patency were observed. So far, no vessel loss has emerged.
CONCLUSION: Conventional repair of PAA has been a standard procedure for many years, though it carries high surgical risk as well as perioperative mortality. Fenestrated endografts may be a promising alternative in selected patients.

Entities:  

Mesh:

Year:  2007        PMID: 17696626     DOI: 10.1177/152660280701400412

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  5 in total

1.  A novel measurement technique for the design of fenestrated stent grafts: Comparison with three-dimensional aorta models.

Authors:  Ji Hoon You; Sung-Gwon Kang; Bong Man Kim
Journal:  Exp Clin Cardiol       Date:  2013

Review 2.  The Current State of Fenestrated and Branched Devices for Abdominal Aortic Aneurysm Repair.

Authors:  Holly L Graves; Benjamin M Jackson
Journal:  Semin Intervent Radiol       Date:  2015-09       Impact factor: 1.513

Review 3.  [Endovascular therapy of para-anastomotic aneurysms of the aorta. Technical options].

Authors:  A Hyhlik-Dürr; M S Bischoff; A S Peters; N Attigah; N Attigha; P Geisbüsch; D Böckler
Journal:  Chirurg       Date:  2013-10       Impact factor: 0.955

4.  Decision uncertainty and value of further research: a case-study in fenestrated endovascular aneurysm repair for complex abdominal aortic aneurysms.

Authors:  Oriana Ciani; David Epstein; Claire Rothery; Rod S Taylor; Mark Sculpher
Journal:  Cost Eff Resour Alloc       Date:  2018-04-16

Review 5.  Endoluminal occlusion devices: technology update.

Authors:  Tobias Zander; Samantha Medina; Guillermo Montes; Lourdes Nuñez-Atahualpa; Michel Valdes; Manuel Maynar
Journal:  Med Devices (Auckl)       Date:  2014-12-01
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.