Literature DB >> 17694384

Distal embolic protection during percutaneous intervention of aorto-coronary venous bypass grafts: the FIRST Trial.

M Lins1, H Heuer, M Haude, P Braun, F Stahl, N Franz, R Simon.   

Abstract

OBJECTIVE: Interventions in aorto-coronary venous bypass grafts (CABG) can cause acute procedural complications due to distal embolization of debris. In the FIRST (First European Investigation Regarding the Systematic use of the TriActiv device) multicenter trial the distal endovascular protection system TriActiv (Kensey Nash) was evaluated during intervention of CABG.
METHODS: 195 patients in 17 centers in Germany with significant disease of a vein graft were enrolled. Inclusion and exclusion criteria were comparable to the SAFER trial.
RESULTS: Acute procedural success was achieved in 98% of cases. Aspirated debris was found in 96.5% of patients. Primary endpoints (MACE at 30 days) occurred in 8.7% of all pts. (ITT). No patient died and 7.2% of patients suffered from MI. The rate of early revascularization was 1.5%. Secondary endpoints (MACCE at 30 days) were found in 9.2% and at hospital discharge in 8.7% of patients.
CONCLUSIONS: The TriActiv system is safe and effective. Normal post procedural flow can be preserved and the MACE rate is with 8.7% considerably low. The FIRST trial supports the growing belief that PCI of CABG should be performed with protection systems.

Entities:  

Mesh:

Year:  2007        PMID: 17694384     DOI: 10.1007/s00392-007-0553-5

Source DB:  PubMed          Journal:  Clin Res Cardiol        ISSN: 1861-0684            Impact factor:   5.460


  13 in total

1.  Creatine kinase-MB enzyme elevation following successful saphenous vein graft intervention is associated with late mortality.

Authors:  M K Hong; R Mehran; G Dangas; G S Mintz; A J Lansky; A D Pichard; K M Kent; L F Satler; G W Stone; M B Leon
Journal:  Circulation       Date:  1999-12-14       Impact factor: 29.690

Review 2.  Recognition of the importance of embolization in atherosclerotic vascular disease.

Authors:  E J Topol; J S Yadav
Journal:  Circulation       Date:  2000-02-08       Impact factor: 29.690

3.  Randomized trial of a distal embolic protection device during percutaneous intervention of saphenous vein aorto-coronary bypass grafts.

Authors:  Donald S Baim; Dennis Wahr; Barry George; Martin B Leon; Joel Greenberg; Donald E Cutlip; Unsal Kaya; Jeffrey J Popma; Kalon K L Ho; Richard E Kuntz
Journal:  Circulation       Date:  2002-03-19       Impact factor: 29.690

4.  Lack of benefit from intravenous platelet glycoprotein IIb/IIIa receptor inhibition as adjunctive treatment for percutaneous interventions of aortocoronary bypass grafts: a pooled analysis of five randomized clinical trials.

Authors:  Marco Roffi; Debabrata Mukherjee; Derek P Chew; Deepak L Bhatt; Leslie Cho; Mark A Robbins; Khaled M Ziada; Danielle M Brennan; Stephen G Ellis; Eric J Topol
Journal:  Circulation       Date:  2002-12-10       Impact factor: 29.690

Review 5.  Distal myocardial protection during percutaneous coronary intervention: when and where?

Authors:  Diana A Gorog; Rodney A Foale; Iqbal Malik
Journal:  J Am Coll Cardiol       Date:  2005-09-29       Impact factor: 24.094

6.  Randomized evaluation of the TriActiv balloon-protection flush and extraction system for the treatment of saphenous vein graft disease.

Authors:  Joseph P Carrozza; Michael Mumma; Jeffrey A Breall; Aland Fernandez; Eugene Heyman; Christopher Metzger
Journal:  J Am Coll Cardiol       Date:  2005-10-10       Impact factor: 24.094

Review 7.  Myonecrosis after revascularization procedures.

Authors:  R M Califf; A E Abdelmeguid; R E Kuntz; J J Popma; C J Davidson; E A Cohen; N S Kleiman; K W Mahaffey; E J Topol; C J Pepine; R J Lipicky; C B Granger; R A Harrington; B E Tardiff; B S Crenshaw; R P Bauman; B D Zuckerman; B R Chaitman; J A Bittl; E M Ohman
Journal:  J Am Coll Cardiol       Date:  1998-02       Impact factor: 24.094

8.  Embolic protection during saphenous vein graft intervention using a second-generation balloon protection device: results from the combined US and European pilot study of the TriActiv Balloon Protected Flush Extraction System.

Authors:  Joseph P Carrozza; Christophe Caussin; Gregory Braden; Peter Braun; Forde Hansell; Robin Fatzinger; Greg Walters; William Kussmaul; Jeffrey Breall
Journal:  Am Heart J       Date:  2005-06       Impact factor: 4.749

9.  Incidence and treatment of 'no-reflow' after percutaneous coronary intervention.

Authors:  R N Piana; G Y Paik; M Moscucci; D J Cohen; C M Gibson; A D Kugelmass; J P Carrozza; R E Kuntz; D S Baim
Journal:  Circulation       Date:  1994-06       Impact factor: 29.690

10.  Local effect of serotonin released during coronary angioplasty.

Authors:  P Golino; F Piscione; C R Benedict; H V Anderson; M Cappelli-Bigazzi; C Indolfi; M Condorelli; M Chiariello; J T Willerson
Journal:  N Engl J Med       Date:  1994-02-24       Impact factor: 91.245

View more
  2 in total

1.  Real-world experience of drug-eluting stents in saphenous vein grafts compared to native coronary arteries: results from the prospective multicenter German DES.DE registry.

Authors:  Ibrahim Akin; Marcus Wiemer; Steffen Schneider; Jochen Senges; Matthias Hochadel; Gert Richardt; Mohamed Abdel-Wahab; Karl-Heinz Kuck; Christoph A Nienaber
Journal:  Clin Res Cardiol       Date:  2011-11-13       Impact factor: 5.460

2.  Predictors and short-term prognosis of angiographically detected distal embolization after emergency percutaneous coronary intervention for ST-elevation acute myocardial infarction.

Authors:  Li Dong-bao; Hua Qi; Liu Zhi; Wang Shan; Jin Wei-ying
Journal:  Clin Res Cardiol       Date:  2009-12       Impact factor: 5.460

  2 in total