Neonatal volume guarantee ventilation: effects of spontaneous breathing, triggered and untriggered inflations.

N McCallion1, R Lau, C J Morley, P A Dargaville.   

Abstract

BACKGROUND: During volume guarantee (VG) ventilation the peak inflating pressure (PIP) for each ventilator inflation is adjusted to ensure the expired tidal volume (V(Te)) is close to the set V(Te). Differences in the PIP between inflations triggered by the infant's inspirations and untriggered inflations are seen. AIM: To investigate the effects of triggered and untriggered inflations on PIP and V(Te).
METHODS: Neonates were ventilated with the Dräger Babylog 8000 using assist control (synchronous intermittent positive pressure ventilation) and VG modes. Continuous recordings of ventilator pressures and tidal volumes were made at 200 Hz for 10 minutes.
RESULTS: In 10 infants, 6540 inflations were analysed, of which 4052 (62%) were triggered. Triggered inflations had a significantly lower mean (SD) PIP than untriggered inflations: 12.9 (4.9) vs 17.0 (3.3) cm H2O, (p<0.001). Despite this, there was no significant difference in the V(Te) of each type of inflation (103% and 101% of the set V(Te), respectively). When a triggered inflation was immediately preceded or followed by an untriggered inflation the PIP changed by about 5 cm H2O. Between adjacent inflations of the same type, the change in PIP was less than 3 cm H2O: for triggered inflations it was 0.11 (1.50) cm H2O and for untriggered inflations 0.06 (1.53) cm H2O.
CONCLUSION: During VG ventilation with the Dräger Babylog 8000 the PIP was 4 cm H2O lower during triggered inflations than untriggered inflations, although the expired tidal volumes were similar.

Entities:  

Mesh:

Year:  2007        PMID: 17686798     DOI: 10.1136/adc.2007.126284

Source DB:  PubMed          Journal:  Arch Dis Child Fetal Neonatal Ed        ISSN: 1359-2998            Impact factor:   5.747


  2 in total

Review 1.  Synchronized mechanical ventilation for respiratory support in newborn infants.

Authors:  Anne Greenough; Thomas E Rossor; Adesh Sundaresan; Vadivelam Murthy; Anthony D Milner
Journal:  Cochrane Database Syst Rev       Date:  2016-09-01

2.  Evaluating peak inspiratory pressures and tidal volume in premature neonates on NAVA ventilation.

Authors:  Alison P Protain; Kimberly S Firestone; Neil L McNinch; Howard M Stein
Journal:  Eur J Pediatr       Date:  2020-07-06       Impact factor: 3.183

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.