Use of the mini C-arm for wrist fractures--establishing a diagnostic reference level.

Gavin J Love1, Anand Pillai, Suzy Gibson.   

Abstract

The establishment of diagnostic reference levels (DRLs) for all typical radiological examinations became mandatory following the implementation of the Ionising Radiations (Medical Exposure) Regulations Act 2000. At present, there are no national dosage guidelines in the UK regarding use of fluoroscopy in orthopaedic trauma. The increasing popularity of the mini C-arm image intensifier amongst surgeons has led to concerns regarding use of ionizing radiation by personnel who have not been trained in radiation protection. It is therefore essential to have formal protocols for use of the mini C-arm to comply with the law and to maintain safe clinical practice. It is attempted to provide dose data for wrist fracture manipulations that may be used as a basis for setting a DRL for this procedure. Screening times were recorded for 80 wrist manipulations in a fracture clinic setting using a mini C-arm image intensifier. A DRL was set using the third quartile value for screening time. The median screening time for wrist fractures was 20 s with a range from 1 to 177 s. The third quartile value for screening time was 34 s. This value can be used as a provisional DRL for wrist fracture manipulations. The DRL is a quantitative guide for the optimisation of radiological protection. IR(ME)R 2000 states that if it is consistently exceeded by an individual operator or a piece of equipment, investigation and remedial action must be taken. We recommend that trauma units establish their own local DRLs for common procedures as made mandatory by legislation.

Entities:  

Mesh:

Year:  2007        PMID: 17681963     DOI: 10.1093/rpd/ncm376

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  1 in total

1.  Establishing local diagnostic reference levels for common orthopaedic procedures using the mini C-arm fluoroscope.

Authors:  Anna Kropelnicki; Rosemary Eaton; Alexandra Adamczyk; Jacqueline Waterman; Pegah Mohaghegh
Journal:  Br J Radiol       Date:  2020-10-22       Impact factor: 3.039

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.