Literature DB >> 17676135

Application of the maximum entropy method to spectral-domain optical coherence tomography for enhancing axial resolution.

Yoshiyuki Takahashi1, Yuuki Watanabe, Manabu Sato.   

Abstract

For the first time we applied the maximum entropy method (MEM) to spectral domain optical coherence tomography to enhance axial resolution (AR). The MEM estimates the power spectrum by fitting. For an onion with optimization of M = 70, the AR of 18.8 microm by discrete Fourier transform (DFT) was improved three times compared with peak widths. The calculation time by the MEM with M = 70 was 20 times longer than that of DFT. However, further studies are needed for practical applications, because the validity of the MEM depends on the sample structures.

Entities:  

Mesh:

Year:  2007        PMID: 17676135     DOI: 10.1364/ao.46.005228

Source DB:  PubMed          Journal:  Appl Opt        ISSN: 1559-128X            Impact factor:   1.980


  1 in total

1.  Optimization for Axial Resolution, Depth Range, and Sensitivity of Spectral Domain Optical Coherence Tomography at 1.3 µm.

Authors:  Sang-Won Lee; Hyun-Woo Jeong; Beop-Min Kim; Yeh-Chan Ahn; Woonggyu Jung; Zhongping Chen
Journal:  J Korean Phys Soc       Date:  2009-12-12       Impact factor: 0.649

  1 in total