No evidence of automatic atrial overdrive pacing efficacy on reduction of paroxysmal atrial fibrillation.

Willem de Voogt1, Norbert van Hemel, Philip de Vusser, Georges H Mairesse, Rob van Mechelen, Juhani Koistinen, Arjan van den Bos, Indrek Roose, Jüri Voitk, Sinikka Yli-Mäyry, Dirk Stockman, Dia El Allaf, Hung-Fat Tse, Chu-Pak Lau.   

Abstract

AIMS: Paroxysmal atrial fibrillation (PAF) is frequently encountered in pacemaker patients, most commonly in sick sinus syndrome. The combination of site-specific pacing in conjunction with an overdrive algorithm combined with antiarrhythmic drugs on the incidence of PAF in patients with a conventional indication for pacing is unknown. METHODS AND
RESULTS: Patients with pacemaker indication and PAF received a DDDR-pacemaker, which included an automatic atrial overdrive (AO) algorithm. The atrial lead was implanted in either the right atrial appendage (RAA) (n = 83) or the right low-atrial septum (LAS) (n = 94). The algorithm was switched on or off in a 3 month, single blind crossover design and antiarrhythmic drugs were kept stable. A control group of 96 patients (LAS, n = 14; RAA, n = 84) without PAF served as controls to assess any proarrhythmic effect of overdrive pacing. Atrial fibrillation (AF) burden defined as cumulative time in mode switch was not reduced during automatic AO from either the RAA or from the LAS. The reduction was not effective both for AF of short (<24 h) and long (> or =24 h) duration. There was no atrial proarrhythmia induced by the overdrive algorithm in the control group.
CONCLUSIONS: We could not demonstrate a reduction of AF burden defined as cumulative time in AF by the AO algorithm, in patients who are paced for standard indications and PAF, neither from the RAA nor from the LAS.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17670782     DOI: 10.1093/europace/eum149

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  4 in total

1.  8 October 1958, D Day for the implantable pacemaker.

Authors:  N M van Hemel; E E van der Wall
Journal:  Neth Heart J       Date:  2008-10       Impact factor: 2.380

2.  Clinical application of pacemakers in atrial tachyarrhythmias.

Authors:  W G de Voogt; N M van Hemel
Journal:  Neth Heart J       Date:  2008-10       Impact factor: 2.380

3.  Diagnostic tools for atrial tachyarrhythmias in implantable pacemakers: a review of technical options and pitfalls.

Authors:  W G de Voogt; N M van Hemel
Journal:  Neth Heart J       Date:  2008-06       Impact factor: 2.380

Review 4.  Advances in Cardiac Pacing: Arrhythmia Prediction, Prevention and Control Strategies.

Authors:  Mehrie Harshad Patel; Shrikanth Sampath; Anoushka Kapoor; Devanshi Narendra Damani; Nikitha Chellapuram; Apurva Bhavana Challa; Manmeet Pal Kaur; Richard D Walton; Stavros Stavrakis; Shivaram P Arunachalam; Kanchan Kulkarni
Journal:  Front Physiol       Date:  2021-12-02       Impact factor: 4.566

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.