Literature DB >> 17670515

Massive CO2 embolism in cardiopulmonary bypass circuit - a near miss.

Dharmendra Agrawal1, Keng-Leong Ang, Sai Prasad.   

Abstract

OBJECTIVES: We report a case of massive air embolism in the cardiopulmonary bypass (CPB) circuit to highlight the principle of its immediate management and its preventive measures.
METHODS: The air embolism was caused by the accidental connection of carbon dioxide (CO(2)) supply to the monitoring arm of the retrograde cardioplegia delivery system. The CPB ceased, resulting in cardiac arrest requiring immediate internal cardiac massage. Cerebral protective measures were also instituted instantaneously and the air embolism was purged from the CPB circuit. The cause of air embolism was identified after high initial arterial CO(2) concentration was detected in the blood gas analysis. The CO(2) supply was disconnected, and CPB was restarted. Spontaneous cardiac activities resumed shortly after, and the operation was completed uneventfully.
RESULTS: The patient had no immediate post-operative sequel and remained well at 6 weeks follow-up.
CONCLUSION: Massive air embolism in cardiopulmonary bypass (CPB) circuit is a life-threatening emergency. Immediate cerebral protective manoeuvres and rectification of the cause of air embolism are vital for favourable outcome. However, high degree of vigilance and cooperation amongst all teams involved are paramount to prevent its occurrence in the first place.

Entities:  

Year:  2005        PMID: 17670515     DOI: 10.1510/icvts.2005.117051

Source DB:  PubMed          Journal:  Interact Cardiovasc Thorac Surg        ISSN: 1569-9285