Literature DB >> 17669889

The way the intra-aortic balloon catheter moves within the aorta as a possible mechanism of balloon associated morbidity.

Haralambos Parissis1, Michael Leotsinidis, Dimitrios Dougenis, David Richens.   

Abstract

This study set off to investigate which mode of weaning of an intra-aortic balloon pump (IABP) produces more aortic trauma. With the use of a perfusion pump, an intact porcine aorta with an IABP in situ, was studied. Angioscopic images of the interior of the aorta were obtained. Whilst keeping steady blood pressure and flow, an 'aortic impact score' was calculated. Endoscopically there is a 'whipping' effect of the balloon shaft on the lateral aortic wall, which appears to be prominent in 1:3 mode. The aortic impact score at 0.5, 6 and 12 h during the experiments was: (1) When weaning by mode: a) 1:1 3.3+/-0.6, 4.0+/-1.0 and 4.3+/-0.6; b) 1:2 4.7+/-0.6, 6.7+/-0.6 and 7.0+/-0.0; c) 1:3 8.7+/-0.6, 11+/-1.0 and 11.7+/-0.6. (2) Weaning by augmentation: a) 75% 2.3+/-0.6, 2.7+/-0.6 and 3.0+/-0.0; b) 50% 1.3+/-0.6, 1.3+/-0.6 and 1.7+/-0.6. An increasing score was observed while weaning by mode (P<0.05). The 1:3 mode produces marked intimal disruption that worsens with time.

Entities:  

Mesh:

Year:  2007        PMID: 17669889     DOI: 10.1510/icvts.2006.147256

Source DB:  PubMed          Journal:  Interact Cardiovasc Thorac Surg        ISSN: 1569-9285


  1 in total

1.  Intra-aortic balloon pumping reduces the increased arterial load caused by acute cardiac depression, modifying central and peripheral load determinants in a time- and flow-related way.

Authors:  Daniel Bia; Edmundo I Cabrera-Fischer; Yanina Zócalo; Ricardo L Armentano
Journal:  Heart Vessels       Date:  2011-11-09       Impact factor: 2.037

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.