Literature DB >> 17665788

Massive femoral osteolytic lesion secondary to disengagement of a total knee tibial locking pin: case report.

Sathappan S Sathappan1, Craig Capeci, Matthew Teicher, Paul E Di Cesare.   

Abstract

The Anatomic Modular Knee (AMK; Depuy, Warsaw, Ind) was one of the first designs to incorporate a modular locking mechanism for the tibial insert; fixation was secured using a screw-in tibial tray-locking pin. This case report describes a patient who presented with instability and worsening knee pain 15 years following a primary total knee arthroplasty using the AMK prosthesis. Radiographs and intraoperative assessment revealed proximal migration of the tibial tray-locking pin into the medial femoral condyle resulting in a large osteolytic defect.

Entities:  

Mesh:

Year:  2007        PMID: 17665788     DOI: 10.1055/s-0030-1248050

Source DB:  PubMed          Journal:  J Knee Surg        ISSN: 1538-8506            Impact factor:   2.757


  1 in total

1.  Failure of tibial polyethylene insert locking mechanism in posterior stabilized arthroplasty.

Authors:  Emmanuel Thienpont
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-05-01       Impact factor: 4.342