Literature DB >> 17664191

The implantable loop recorder in children.

B Yeung1, K McLeod.   

Abstract

OBJECTIVE: To evaluate use of the implantable loop recorder in children.
SETTING: Royal Hospital for Sick Children, Glasgow, UK.
METHOD: Retrospective study of children who had an implantable loop recorder between September 1998 and October 2005.
RESULTS: 38 devices were implanted in 34 children. Median age at implantation 11.3 years (range 1.8-17.6); median follow-up period 15 months (range 2-58). The main presenting complaint was syncope in 26 (76.5%), seizures in 6 (17.6%) and palpitations in 2 (5.9%). After implantation, 19 (55.9%) patients had symptom recurrence. Of these, 11 were shown to have sinus rhythm during symptoms and 8 had an abnormal ECG. Four patients had asystole >3 seconds and were diagnosed with reflex asystolic syncope; 2 had polymorphic ventricular tachycardia. One patient who already had a diagnosis of long QT syndrome was shown to have ventricular ectopy during symptoms and beta-blockers were increased. One patient had transient complete heart block during symptoms but refused a pacemaker. In almost half the patients (44.1%), symptoms resolved after implantation. Complications requiring removal of the device occurred in 6 (15.8%) implants.
CONCLUSIONS: In children with syncope and palpitations, the implantable loop recorder appears to be an excellent method of effecting a "cure" in almost 50% of subjects. For those who remain symptomatic, it is successful in determining cardiac rhythm during symptoms, but the complication rate in children may be higher than that of adults.

Entities:  

Mesh:

Year:  2007        PMID: 17664191     DOI: 10.1136/hrt.2007.121855

Source DB:  PubMed          Journal:  Heart        ISSN: 1355-6037            Impact factor:   5.994


  1 in total

1.  Idiopathic ventricular arrhythmias detected by an implantable loop recorder in a child with exercise-induced syncope.

Authors:  Celal Akdeniz; Isa Ozyilmaz; Murat Saygi; Yakup Ergul; Volkan Tuzcu
Journal:  Tex Heart Inst J       Date:  2013
  1 in total