Esophageal perforation from anterior cervical screw migration.

Ramesh L Sahjpaul1.   

Abstract

BACKGROUND: Esophageal perforation from anterior cervical instrumentation migration is an uncommon but potentially highly morbid or even fatal complication. Early recognition and aggressive investigation and treatment are essential to ensure good outcome. CASE DESCRIPTION: A 58-year-old man underwent C6 vertebrectomy and C5-7 interbody fusion with a cage and anterior cervical plate. After surgery he developed fever and recurrence of his symptoms and deficits, but was managed expectantly. He was then referred to the author's institution. A barium swallow demonstrated an esophageal fistula (a Gastrograffin swallow was falsely negative) caused by a migrated screw; serial radiographs confirmed its passage through the gastrointestinal tract. Revision surgery was required to repair the perforation and reconstruct the cervicothoracic spine. Intraoperative esophageal injection of methylene blue was helpful in demonstrating the site of leakage. Despite a prolonged postoperative course complicated by pulmonary embolus, the patient recovered with minimal residual deficit, and continues to do well 2 years later.
CONCLUSIONS: A high index of suspicion followed by aggressive investigation are crucial in the setting of unexpected neck pain, new neurologic deficit, fever, or swallowing difficulties in the early postoperative period after anterior cervical spine instrumentation. If esophageal perforation is suspected, a barium swallow is recommended over Gastrograffin, which, although less irritating to the surrounding tissues, may be falsely negative. Intraoperative methylene blue injection into the esophageal lumen is useful in identifying the site of perforation.

Entities:  

Mesh:

Year:  2007        PMID: 17662361     DOI: 10.1016/j.surneu.2006.09.032

Source DB:  PubMed          Journal:  Surg Neurol        ISSN: 0090-3019


  19 in total

1.  Early Radiographic and Clinical Outcomes Study Evaluating an Integrated Screw and Interbody Spacer for One- and Two-Level ACDF.

Authors:  Paul D Lane; Jacob L Cox; Roger B Gaskins; Brandon G Santoni; James B Billys; Antonio E Castellvi
Journal:  Int J Spine Surg       Date:  2015-07-17

2.  Erosion of anterior cervical plate into pharynx with pharyngotracheal fistula.

Authors:  Yuo-Chen Kuo; Marc S Levine
Journal:  Dysphagia       Date:  2010-01-21       Impact factor: 3.438

3.  Anterior cervical interbody fusion with the Zero-P spacer: mid-term results of two-level fusion.

Authors:  Yuanyuan Chen; Huajing Chen; Peng Cao; Wen Yuan
Journal:  Eur Spine J       Date:  2015-04-08       Impact factor: 3.134

4.  Comparison of Radiologic Outcomes of Different Methods in Single-Level Anterior Cervical Discectomy and Fusion.

Authors:  O Ik Kwon; Dong Wuk Son; Sang Weon Lee; Geun Sung Song
Journal:  Korean J Spine       Date:  2016-09-30

5.  Rod migration into the posterior fossa after harms operation : case report and review of literatures.

Authors:  Hyoung-Joon Chun; Koang Hum Bak; Tae Hoon Kang; Hyeong-Joong Yi
Journal:  J Korean Neurosurg Soc       Date:  2010-03-31

6.  The application of zero-profile anchored spacer in anterior cervical discectomy and fusion.

Authors:  Zhiwen Wang; Weimin Jiang; Xuefeng Li; Heng Wang; Jinhui Shi; Jie Chen; Bin Meng; Huilin Yang
Journal:  Eur Spine J       Date:  2014-10-22       Impact factor: 3.134

7.  Recurrent esophageal perforation after anterior cervical spine surgery: case report.

Authors:  Man-Kyu Park; Dae-Chul Cho; Woo-Seok Bang; Kyoung-Tae Kim; Joo-Kyung Sung
Journal:  Eur Spine J       Date:  2018-03-02       Impact factor: 3.134

8.  A new zero-profile implant for stand-alone anterior cervical interbody fusion.

Authors:  M Scholz; K J Schnake; A Pingel; R Hoffmann; F Kandziora
Journal:  Clin Orthop Relat Res       Date:  2011-03       Impact factor: 4.176

Review 9.  Zero-P: a new zero-profile cage-plate device for single and multilevel ACDF. A single institution series with four years maximum follow-up and review of the literature on zero-profile devices.

Authors:  Giuseppe M V Barbagallo; Dario Romano; Francesco Certo; Pietro Milone; Vincenzo Albanese
Journal:  Eur Spine J       Date:  2013-09-24       Impact factor: 3.134

Review 10.  Esophageal perforation after anterior cervical surgery: a review of the literature for over half a century with a demonstrative case and a proposed novel algorithm.

Authors:  Ferhat Harman; Erkan Kaptanoglu; Askin Esen Hasturk
Journal:  Eur Spine J       Date:  2016-02-02       Impact factor: 3.134

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.