Literature DB >> 1764837

Relation of mean pixel intensity to concentration of sonicated albumin microspheres: effects of ultrasound system settings.

R A Wray1, M A Quiñones, W A Zoghbi, J Cheirif.   

Abstract

Myocardial contrast echocardiography (MCE) is a recently developed technique to assess myocardial perfusion. Previous studies have shown that MCE frequently underestimates coronary reserve as compared with other methods. Previous reports have suggested that some of the limitations of MCE are due to the nonlinear processing algorithms of the ultrasound systems. Therefore we designed an in vitro model to study the effects of various processing algorithms and levels of acoustic power (AP) on the relation between concentration of echocontrast and mean pixel intensities. A beaker containing a commercially produced, diluted solution of sonicated albumin microspheres was imaged with a commercial Hewlett Packard ultrasound system, and mean pixel intensities were determined with an off-line computer system. The solution was imaged at a wide range of APs (40-10 dB) and time gain compensation (40-60 dB). Concentration versus intensity curves were generated using all possible combinations of compression and processing algorithms available on the ultrasound system used. Contrast effect diminished rapidly when exposed to high AP. Lower powers had less effect on contrast intensity duration, with no effect seen below 20 dB. Changes in time gain compensation did not affect contrast intensities. Regardless of the processing algorithm examined, the relation between concentration of echocontrast and mean pixel intensities was near linear in the first two to three concentrations; thereafter, the curves flattened during increasing concentrations of contrast. At higher concentrations, acoustic shadowing produced a decrease in pixel intensities.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1764837     DOI: 10.1002/clc.4960141706

Source DB:  PubMed          Journal:  Clin Cardiol        ISSN: 0160-9289            Impact factor:   2.882


  1 in total

1.  Abnormal myocardial perfusion and contractile recruitment during exercise in type 1 diabetic patients.

Authors:  Roldano Scognamiglio; Christian Negut; Saula Vigili de Kreuizenberg; Monica Palisi; Antonio Tiengo; Angelo Avogaro
Journal:  Clin Cardiol       Date:  2005-02       Impact factor: 2.882

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.