A newly developed perfused umbrella electrode for radiofrequency ablation: an ex vivo evaluation study in bovine liver.

Philipp Bruners1, Jochen Pfeffer, Rana M Kazim, Rolf W Günther, Thomas Schmitz-Rode, Andreas H Mahnken.   

Abstract

The purpose of this study was to evaluate the effectiveness of a newly developed perfused monopolar radiofrequency (RF) probe with an umbrella-shaped array. A perfused umbrella-shaped monopolar RF probe based on a LeVeen electrode (Boston Scientific Corp., Natick, MA, USA) with a 3-cm array diameter was developed. Five different configurations of this electrode were tested: (a) perfusion channel/endhole, (b) perfusion channel/endhole + sideholes, (c) 1 cm insulation removed at the tip, (d) 1 cm insulation removed at the tip + perfusion channel/endhole, and (e) 1 cm insulation removed at the tip + perfusion channel/endhole + sideholes. An unmodified LeVeen electrode served as a reference standard. RF ablations were performed in freshly excised bovine liver using a commercial monopolar RF system with a 200-W generator (RF 3000; Boston Scientific Corp.). Each electrode was tested 10 times applying the vendor's recommended ablation protocol combined with the preinjection of 2 ml 0.9% saline. Volumes and shapes of the lesions were compared. Lesions generated with the original LeVeen electrode showed a mean volume of 12.74 +/- 0.52 cm(3). Removing parts of the insulation led to larger coagulation volumes (22.65 +/- 2.12 cm(3)). Depending on the configuration, saline preinjection resulted in a further increase in coagulation volume (25.22 +/- 3.37 to 31.28 +/- 2.32 cm(3)). Besides lesion volume, the shape of the ablation zone was influenced by the configuration of the electrode used. We conclude that saline preinjection in combination with increasing the active tip length of the umbrella-shaped LeVeen RF probe allows the reliable ablation of larger volumes in comparison to the originally configured electrode.

Entities:  

Mesh:

Year:  2007        PMID: 17647060     DOI: 10.1007/s00270-007-9125-3

Source DB:  PubMed          Journal:  Cardiovasc Intervent Radiol        ISSN: 0174-1551            Impact factor:   2.740


  1 in total

1.  Perfused hypertonic-saline-augmented needle enlarges ablation zones in ex vivo porcine livers.

Authors:  Caoye Wang; Qi Wang; Dachuan Zhang; Zhongming He; Wenhua Chen
Journal:  Oncol Lett       Date:  2017-11-06       Impact factor: 2.967

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.