Improving diagnostic ability of blood oxygen saturation from overnight pulse oximetry in obstructive sleep apnea detection by means of central tendency measure.

Daniel Alvarez1, Roberto Hornero, María García, Félix del Campo, Carlos Zamarrón.   

Abstract

OBJECTIVES: Nocturnal pulse oximetry is a widely used alternative to polysomnography (PSG) in screening for obstructive sleep apnea (OSA) syndrome. Several oximetric indexes have been derived from nocturnal blood oxygen saturation (SaO2). However, they suffer from several limitations. The present study is focused on the usefulness of nonlinear methods in deriving new measures from oximetry signals to improve the diagnostic accuracy of classical oximetric indexes. Specifically, we assessed the validity of central tendency measure (CTM) as a screening test for OSA in patients clinically suspected of suffering from this disease.
MATERIALS AND METHODS: We studied 187 subjects suspected of suffering from OSA referred to the sleep unit. A nocturnal pulse oximetry study was applied simultaneously to a conventional PSG. Three different index groups were compared. The first one was composed by classical indexes provided by our oximeter: oxygen desaturation indexes (ODIs) and cumulative time spent below a saturation of 90% (CT90). The second one was formed by indexes derived from a nonlinear method previously studied by our group: approximate entropy (ApEn). The last one was composed by indexes derived from a CTM analysis.
RESULTS: For a radius in the scatter plot equal to 1, CTM values corresponding to OSA positive patients (0.30+/-0.20, mean+/-S.D.) were significantly lower (p<<0.001) than those values from OSA negative subjects (0.71+/-0.18, mean+/-S.D.). CTM was significantly correlated with classical indexes and indexes from ApEn analysis. CTM provided the highest correlation with the apnea-hipopnea index AHI (r=-0.74, p<0.0001). Moreover, it reached the best results from the receiver operating characteristics (ROC) curve analysis, with 90.1% sensitivity, 82.9% specificity, 88.5% positive predictive value, 85.1% negative predictive value, 87.2% accuracy and an area under the ROC curve of 0.924. Finally, the AHI derived from the quadratic regression curve for the CTM showed better agreement with the AHI from PSG than classical and ApEn derived indexes.
CONCLUSION: The results suggest that CTM could improve the diagnostic ability of SaO2 signals recorded from portable monitoring. CTM could be a useful tool for physicians in the diagnosis of OSA syndrome.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17643971     DOI: 10.1016/j.artmed.2007.06.002

Source DB:  PubMed          Journal:  Artif Intell Med        ISSN: 0933-3657            Impact factor:   5.326


  17 in total

1.  Oximetry as an Accurate Tool for Identifying Moderate to Severe Sleep Apnea in Patients With Acute Stroke.

Authors:  Shih Hao Lin; Chantale Branson; Jamie Leung; Lisa Park; Nirmita Doshi; Sanford H Auerbach
Journal:  J Clin Sleep Med       Date:  2018-12-15       Impact factor: 4.062

2.  Assessment of heart rate variability by application of central tendency measure.

Authors:  Laurita dos Santos; Joaquim J Barroso; Elbert E N Macau; Moacir F de Godoy
Journal:  Med Biol Eng Comput       Date:  2015-09-22       Impact factor: 2.602

3.  Novel mathematical processing method of nocturnal oximetry for screening patients with suspected sleep apnoea syndrome.

Authors:  Laurent Poupard; Carole Philippe; Michael David Goldman; Richard Sartène; Marc Mathieu
Journal:  Sleep Breath       Date:  2011-04-15       Impact factor: 2.816

4.  Nocturnal Oximetry-based Evaluation of Habitually Snoring Children.

Authors:  Roberto Hornero; Leila Kheirandish-Gozal; Gonzalo C Gutiérrez-Tobal; Mona F Philby; María Luz Alonso-Álvarez; Daniel Álvarez; Ehab A Dayyat; Zhifei Xu; Yu-Shu Huang; Maximiliano Tamae Kakazu; Albert M Li; Annelies Van Eyck; Pablo E Brockmann; Zarmina Ehsan; Narong Simakajornboon; Athanasios G Kaditis; Fernando Vaquerizo-Villar; Andrea Crespo Sedano; Oscar Sans Capdevila; Magnus von Lukowicz; Joaquín Terán-Santos; Félix Del Campo; Christian F Poets; Rosario Ferreira; Katalina Bertran; Yamei Zhang; John Schuen; Stijn Verhulst; David Gozal
Journal:  Am J Respir Crit Care Med       Date:  2017-12-15       Impact factor: 21.405

5.  Artificial apnea classification with quantitative sleep EEG synchronization.

Authors:  Mehmet Akṣahin; Serap Aydın; Hikmet Fırat; Osman Eroǧul
Journal:  J Med Syst       Date:  2010-03-16       Impact factor: 4.460

6.  Automated Screening of Children With Obstructive Sleep Apnea Using Nocturnal Oximetry: An Alternative to Respiratory Polygraphy in Unattended Settings.

Authors:  Daniel Álvarez; María L Alonso-Álvarez; Gonzalo C Gutiérrez-Tobal; Andrea Crespo; Leila Kheirandish-Gozal; Roberto Hornero; David Gozal; Joaquín Terán-Santos; Félix Del Campo
Journal:  J Clin Sleep Med       Date:  2017-05-15       Impact factor: 4.062

7.  Detection of sleep disordered breathing and its central/obstructive character using nasal cannula and finger pulse oximeter.

Authors:  Dirk Sommermeyer; Ding Zou; Ludger Grote; Jan Hedner
Journal:  J Clin Sleep Med       Date:  2012-10-15       Impact factor: 4.062

Review 8.  Computer-Assisted Diagnosis of the Sleep Apnea-Hypopnea Syndrome: A Review.

Authors:  Diego Alvarez-Estevez; Vicente Moret-Bonillo
Journal:  Sleep Disord       Date:  2015-07-21

9.  The role of overnight pulse-oximetry in recognition of obstructive sleep apnea syndrome in morbidly obese and non obese patients.

Authors:  Stefan Dumitrache-Rujinski; George Calcaianu; Dragos Zaharia; Claudia Lucia Toma; Miron Bogdan
Journal:  Maedica (Buchar)       Date:  2013-09

10.  Early diagnosis of sleep related breathing disorders.

Authors:  Joachim T Maurer
Journal:  GMS Curr Top Otorhinolaryngol Head Neck Surg       Date:  2010-10-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.