An empirical characterization of the quality of DTI data and the efficacy of dyadic sorting.

Nathan E Yanasak1, Jerry D Allison, Tom C-C Hu.   

Abstract

Metrics calculated from images acquired using the diffusion tensor imaging (DTI) technique possess a systematic bias that depends on signal-to-noise ratio (SNR). Dyadic sorting provides a simple method for remediating some of this bias within a region(s) of interest (ROI). Although this bias and its removal using dyadic sorting have been studied previously within a theoretical framework, one can employ precise geometric knowledge of microstructures to perform an empirical comparison between expected DTI results and those measured with a scanner. In this project, the biasing effect of low SNR (approximately 1-10) on DTI eigenvalues was measured directly using water-filled capillary structures of two different sizes, and the magnitude of the corrective effect of dyadically sorting eigenvector-eigenvalue pairs was characterized. Multiple DTI series were acquired for determining DTI metrics at eight unique SNR values, using T(R) to vary signal intensity via T(1) contrast. Differences between the second and third eigenvalues, which should be equal for prolate geometry, ranged from approximately 23% to 45% and from 19% to 41% for large and small inner diameter capillaries after sorting eigenvalues by magnitude, and ranged from approximately 1% to 18% and from 1% to 4% after dyadic sorting. A high-resolution DTI series was used to observe the effect of ROI size on dyadic sorting. For restriction of diffusion on the scale of the small capillary at SNR approximately 18, an ROI with > or =50 pixels is adequate to determine fractional anisotropy to 99% accuracy, while larger ROI are required to resolve the two smaller eigenvalues to the same accuracy ( approximately 330-390 pixels). At low values of SNR, the iteration of dyadic sorting is suggested to achieve good accuracy. A method for the incorporation of empirical measurements into a bias-correction map, which would be useful for characterizing uncertainty and for reducing systematic bias in DTI data, is introduced.

Entities:  

Mesh:

Year:  2007        PMID: 17643898     DOI: 10.1016/j.mri.2007.05.006

Source DB:  PubMed          Journal:  Magn Reson Imaging        ISSN: 0730-725X            Impact factor:   2.546


  3 in total

Review 1.  Multicentre imaging measurements for oncology and in the brain.

Authors:  P S Tofts; D J Collins
Journal:  Br J Radiol       Date:  2011-12       Impact factor: 3.039

2.  Quantification of accuracy and precision of multi-center DTI measurements: a diffusion phantom and human brain study.

Authors:  Tong Zhu; Rui Hu; Xing Qiu; Michael Taylor; Yuen Tso; Constantin Yiannoutsos; Bradford Navia; Susumu Mori; Sven Ekholm; Giovanni Schifitto; Jianhui Zhong
Journal:  Neuroimage       Date:  2011-02-18       Impact factor: 6.556

3.  Comparison of Glass Capillary Plates and Polyethylene Fiber Bundles as Phantoms to Assess the Quality of Diffusion Tensor Imaging.

Authors:  Atsushi Tachibana; Yasuhiko Tachibana; Jeff Kershaw; Hiromi Sano; Masahiro Fukushi; Takayuki Obata
Journal:  Magn Reson Med Sci       Date:  2017-12-05       Impact factor: 2.471

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.