Development of automated amperometric detection of antibodies against Bacillus anthracis protective antigen.

Zoraida P Aguilar1, Mali Sirisena.   

Abstract

Picogram levels of antibodies against the protective antigen (PA) of Bacillus anthracis were detected in an automated electrochemical sandwich-type enzyme-linked immunosorbant assay. The antibodies were captured and detected using an 8 x 3 array of 50-microm-diameter cavities. The reagent and sample volumes were as low as 200 nL in a less than 25-min assay from capture to signal generation. The electrochemical detection of the antibodies was demonstrated at 0.05-10 microg/mL containing only 10-5,000 pg antibodies. The limit of detection is 10 fg for a 200-nL sample. Detection of anti-PA immunoglobulin G performed in spiked normal human serum and fresh whole human blood did not show a significant difference from detection in a buffer. The initial automation of the assay involved the use of a digital syringe pump for the delivery of reagents to the capture surface.