Literature DB >> 17634658

Dose integration characteristics in normoxic polymer gel dosimetry investigated using sequential beam irradiation.

A Karlsson1, H Gustavsson, S Månsson, K B McAuley, S A J Bäck.   

Abstract

Dose integration properties were investigated for normoxic polymer gels based on methacrylic acid (nMAG) and acrylamide/N, N'-methylenebisacrylamide (nPAG). The effect of sequential irradiation was studied for different fractionation schemes and varying amounts of methacrylic acid for the nMAG gels. Magnetic resonance imaging (MRI) was used for read out of the absorbed dose response. The investigated gels exhibited a dependence on the fractionation scheme. The response when the total dose was divided into fractions of 0.5 Gy was compared with the response when the total dose was delivered in a single fraction. The slope of the R2 versus the absorbed dose response decreased when the absorbed dose per fraction was increased. Also, for higher amounts of methacrylic acid in the nMAG system the difference in the response increased. For gels containing 2, 4, 6 and 8% methacrylic acid, the R2 versus the absorbed dose response increased by 35, 37, 63 and 93%, respectively. Furthermore, the effect of the fractionation was larger when a higher total absorbed dose was given. The effect was less pronounced for the investigated nPAG, containing 3% acrylamide and 3% N, N'-methylenebisacrylamide, than for the nMAG systems. Consequently, this study indicates that the nPAG system has preferable beam integration characteristics compared with the nMAG system.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17634658     DOI: 10.1088/0031-9155/52/15/021

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  11 in total

1.  Improved MAGIC gel for higher sensitivity and elemental tissue equivalent 3D dosimetry.

Authors:  Xuping Zhu; Timothy G Reese; Elizabeth M Crowley; Georges El Fakhri
Journal:  Med Phys       Date:  2010-01       Impact factor: 4.071

Review 2.  Polymer gel dosimetry.

Authors:  C Baldock; Y De Deene; S Doran; G Ibbott; A Jirasek; M Lepage; K B McAuley; M Oldham; L J Schreiner
Journal:  Phys Med Biol       Date:  2010-02-11       Impact factor: 3.609

3.  Influence of magnesium chloride on the dose-response of polyacrylamide-type gel dosimeters.

Authors:  Shin-Ichiro Hayashi; Hiraku Kawamura; Shuji Usui; Takahiro Tominaga
Journal:  Radiol Phys Technol       Date:  2018-09-04

4.  Application of radiochromic gel dosimetry to commissioning of a megavoltage research linear accelerator for small-field animal irradiation studies.

Authors:  Noora Ba Sunbul; Ibrahim Oraiqat; Benjamin Rosen; Cameron Miller; Christopher Meert; Martha M Matuszak; Shaun Clarke; Sara Pozzi; Jean M Moran; Issam El Naqa
Journal:  Med Phys       Date:  2021-02-06       Impact factor: 4.071

5.  Methylphenidate for attention deficit hyperactivity disorder and drug relapse in criminal offenders with substance dependence: a 24-week randomized placebo-controlled trial.

Authors:  Maija Konstenius; Nitya Jayaram-Lindström; Joar Guterstam; Olof Beck; Björn Philips; Johan Franck
Journal:  Addiction       Date:  2013-12-01       Impact factor: 6.526

6.  Determine the Dose Distribution Using Ultrasound Parameters in MAGIC-f Polymer Gels.

Authors:  Hossein Masoumi; Manijhe Mokhtari-Dizaji; Azim Arbabi; Mohsen Bakhshandeh
Journal:  Dose Response       Date:  2016-02-11       Impact factor: 2.658

7.  Dose and slice thickness evaluation with nMAG gel dosimeters in computed tomography.

Authors:  Chun-Chao Chuang; Jay Wu
Journal:  Sci Rep       Date:  2018-02-08       Impact factor: 4.379

8.  Basic Properties of a New Polymer Gel for 3D-Dosimetry at High Dose-Rates Typical for FFF Irradiation Based on Dithiothreitol and Methacrylic Acid (MAGADIT): Sensitivity, Range, Reproducibility, Accuracy, Dose Rate Effect and Impact of Oxygen Scavenger.

Authors:  Muzafar Khan; Gerd Heilemann; Wolfgang Lechner; Dietmar Georg; Andreas Georg Berg
Journal:  Polymers (Basel)       Date:  2019-10-19       Impact factor: 4.329

9.  Application of 3D Gel Dosimetry as a Quality Assurance Tool in Functional Leksell Gamma Knife Radiosurgery.

Authors:  Linas Kudrevicius; Evelina Jaselske; Diana Adliene; Viktoras Rudzianskas; Andrius Radziunas; Arimantas Tamasauskas
Journal:  Gels       Date:  2022-01-20

10.  Small-Field Measurements of 3D Polymer Gel Dosimeters through Optical Computed Tomography.

Authors:  Tian-Yu Shih; Jay Wu; Cheng-Ting Shih; Yao-Ting Lee; Shin-Hua Wu; Chun-Hsu Yao; Bor-Tsung Hsieh
Journal:  PLoS One       Date:  2016-03-14       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.