Literature DB >> 17630716

Measurement of mobile-phase volume in reversed-phase liquid chromatography and evaluation of the composition of liquid layer formed by solvation of packing materials.

Masami Shibukawa1, Yuji Takazawa, Kazunori Saitoh.   

Abstract

The mobile-phase volumes (Vm) in reversed-phase liquid chromatography (RPLC) with alkyl-bonded silica, defined as the difference between the total volume of eluent in the column (V0) and the volume of the eluent solvent layer formed by solvation of the bonded phase (VL), are determined by the method derived from the eluent electrolyte effect on the retention of ionic analytes. The validity of the Vm values obtained is evaluated by comparing them with the retention volumes of various organic compounds and inorganic ions, which have been suggested as unretained markers, and those obtained from a linear dependence of the logarithmic retention factor on the carbon numbers of homologous series. From the results obtained, it has been concluded that the solvated liquid phase on a column packing material should be assigned to a part of the stationary phase and the method developed for determination of the Vm value based on the ion partition model gives the most reasonable value as the mobile-phase volume in RPLC. The volume and the solvent composition of the solvated liquid phase on C1, C8, and C18 bonded silica are estimated, and the effects of organic modifiers and the physicochemical structures of the packing materials on these values are discussed.

Entities:  

Year:  2007        PMID: 17630716     DOI: 10.1021/ac0701839

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  1 in total

1.  Characterization of the Interfacial Liquid Layer Formed on Hydrophobic Packing Material Surfaces by Liquid Chromatographic Analysis of the Distribution of Ions and Molecules.

Authors:  Masami Shibukawa; Hiroki Okutsu; Shingo Saito
Journal:  ACS Omega       Date:  2022-04-19
  1 in total