Eisuke Hanada1, Shuuya Itoga, Kyoko Takano, Takato Kudou.
Abstract
Medical devices driven by electric power have come to be commonly used in hospitals, and rapid changes of voltage or current can easily cause them to fail. A stable and high quality power supply is indispensable in order to maintain safety in the modern clinical setting. Therefore, we investigated the quality of the power supply in a hospital and determined the tolerance of 13 pieces of medical equipment to voltage dips. The results showed little distortion of the voltage wave. However, we found an approximately 7% momentary voltage dip caused by lightening and other problems, such as 2 to 5% periodic drops in voltage and voltage wave distortions caused by incorrect grounding. In a tolerance test, the settings of some medical devices were changed at the time of automatic reboot after a disturbance. For another device, trend information was initialized.Mesh:
Year: 2007 PMID: 17622025 DOI: 10.1007/s10916-007-9058-z