Brain tumor enhancement in MR imaging at 3 Tesla: comparison of SNR and CNR gain using TSE and GRE techniques.

Bernd J Wintersperger1, Val M Runge, Jonmenjoy Biswas, Maximilian F Reiser, Stefan O Schoenberg.   

Abstract

PURPOSE: The purpose of this study was to compare brain and tumor signal characteristics of T1-weighted turbo spin-echo (TSE) and gradient recalled echo (GRE) sequence techniques at 3 T compared to TSE at 1.5 T, focusing on the detection of contrast enhancement, in a standardized animal model of a brain glioma.
MATERIALS AND METHODS: Twelve rats with implanted brain gliomas were evaluated at 1.5 and 3 T using matched hardware configurations. At 1.5 T, scanning was performed using a TSE sequence optimized for field strength (480/15 milliseconds; 125 Hz/Px) with postcontrast scans acquired at multiple time points after gadoteridol injection (0.1 mmol/kg). At 3 T, scanning was performed using the 1.5 T equivalent TSE as well as with TSE and GRE techniques optimized for 3 T. Signal-to-noise ratio (SNR) of brain and tumor and tumor contrast-to-noise ratio (CNR) were evaluated for all techniques at both field strengths.
RESULTS: Postcontrast tumor SNR (63.7 +/- 10.8 vs. 29.5 +/- 4.3; P < 0.0001) and brain SNR (35.8 +/- 1.5 vs. 19.1 +/- 0.7; P < 0.0001) showed significant increase at 3 T using matched TSE. Comparing TSE optimized to each field strength (for optimized gray-white contrast), tumor and brain SNR still showed a significant increase at 3 T of 73% and 56%, respectively (both P < 0.0001). Comparing TSE at 1.5 T and GRE at 3 T, tumor SNR increased by 105%, whereas brain SNR increased by 141% (both P < 0.0001). Tumor CNR with matched TSE increased by 168% (P < 0.0001), with optimized TSE by 111% (P < 0.0001), and with GRE at 3 T versus TSE at 1.5 T by 36% (P < 0.001). With additional adjustments for echo time the gain in tumor CNR for 2D GRE may again reach 60%.
CONCLUSIONS: With TSE at 3 T, the SNR gain comes close to the theoretically expected doubling with an even higher tumor CNR increase. In a clinical like setting at 3 T, where a T1w GRE sequence is used, tumor CNR gain is limited. Contrast dose should therefore not be decreased at 3 T.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17620938     DOI: 10.1097/RLI.0b013e31803e8b3f

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  12 in total

Review 1.  Consensus recommendations for a standardized Brain Tumor Imaging Protocol in clinical trials.

Authors:  Benjamin M Ellingson; Martin Bendszus; Jerrold Boxerman; Daniel Barboriak; Bradley J Erickson; Marion Smits; Sarah J Nelson; Elizabeth Gerstner; Brian Alexander; Gregory Goldmacher; Wolfgang Wick; Michael Vogelbaum; Michael Weller; Evanthia Galanis; Jayashree Kalpathy-Cramer; Lalitha Shankar; Paula Jacobs; Whitney B Pope; Dewen Yang; Caroline Chung; Michael V Knopp; Soonme Cha; Martin J van den Bent; Susan Chang; W K Al Yung; Timothy F Cloughesy; Patrick Y Wen; Mark R Gilbert
Journal:  Neuro Oncol       Date:  2015-08-05       Impact factor: 12.300

Review 2.  MR imaging of neoplastic central nervous system lesions: review and recommendations for current practice.

Authors:  M Essig; N Anzalone; S E Combs; À Dörfler; S-K Lee; P Picozzi; A Rovira; M Weller; M Law
Journal:  AJNR Am J Neuroradiol       Date:  2011-10-20       Impact factor: 3.825

3.  Comparing 3T T1-weighted sequences in identifying hyperintense punctate lesions in preterm neonates.

Authors:  D Tortora; V Panara; P A Mattei; A Tartaro; R Salomone; S Domizio; A R Cotroneo; M Caulo
Journal:  AJNR Am J Neuroradiol       Date:  2014-11-06       Impact factor: 3.825

Review 4.  Radiogenomics and imaging phenotypes in glioblastoma: novel observations and correlation with molecular characteristics.

Authors:  Benjamin M Ellingson
Journal:  Curr Neurol Neurosci Rep       Date:  2015-01       Impact factor: 5.081

5.  Emerging techniques and technologies in brain tumor imaging.

Authors:  Benjamin M Ellingson; Martin Bendszus; A Gregory Sorensen; Whitney B Pope
Journal:  Neuro Oncol       Date:  2014-10       Impact factor: 12.300

6.  Use of a clinical MRI scanner for preclinical research on rats.

Authors:  Akihide Yamamoto; Hiroshi Sato; Jun-ichiro Enmi; Kenji Ishida; Takayuki Ose; Atsuomi Kimura; Hideaki Fujiwara; Hiroshi Watabe; Takuya Hayashi; Hidehiro Iida
Journal:  Radiol Phys Technol       Date:  2008-10-01

7.  Comparison of spin-echo T1- and T2-weighted and gradient-echo T1-weighted images at 3T in evaluating very preterm neonates at term-equivalent age.

Authors:  B Sarikaya; A M McKinney; B Spilseth; C L Truwit
Journal:  AJNR Am J Neuroradiol       Date:  2012-12-06       Impact factor: 3.825

8.  MRI assessment of experimental gliomas using 17.6 T.

Authors:  Marc A Schwarz; Mirko Pham; Xavier Helluy; Arnd Doerfler; Tobias Engelhorn
Journal:  Neuroradiology       Date:  2013-03-10       Impact factor: 2.804

9.  Comparison of spin-echo and gradient-echo T1-weighted and spin-echo T2-weighted images at 3T in evaluating term-neonatal myelination.

Authors:  A E Tyan; A M McKinney; T J Hanson; C L Truwit
Journal:  AJNR Am J Neuroradiol       Date:  2014-09-11       Impact factor: 3.825

10.  Cerebral staging of lung cancer: is one single contrast-enhanced T1-weighted three-dimensional gradient-echo sequence sufficient?

Authors:  Mickaël Ohana; Mi-Young Jeung; Gauthier Bazille; Catherine Roy
Journal:  Neuroradiology       Date:  2014-05-15       Impact factor: 2.804

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.