An evaluation of the performance of Sysmex XE-2100 in enumerating nucleated red cells in peripheral blood.

Gene Gulati1, Eric Behling, William Kocher, Roland Schwarting.   

Abstract

CONTEXT: Automated methods of enumerating nucleated red blood cells (NRBCs) in blood are gaining acceptance in many laboratories.
OBJECTIVE: To evaluate the performance of Sysmex XE-2100 in enumerating NRBCs in peripheral blood.
DESIGN: Automated relative number of NRBCs per 100 white blood cells (NRBC%) results for a total of 460 specimens run on the XE-2100 were compared with manual NRBC% results obtained by performing a 100-cell differential on a Wright-stained smear prepared from each specimen. To assess within-day reproducibility, 64 specimens were rerun on the XE-2100, and a second 100-cell differential was performed on the original blood smear. Excel software was used for data analysis.
RESULTS: Regression analysis of automated NRBC% versus manual NRBC% yielded a correlation coefficient of 0.9712. No NRBCs were seen in the blood smear on 35 (15.1%) of 232 specimens with automated NRBC% of 0.1 to 1.9. The XE-2100 generated an NRBC% of 0.0 on 5 (6.8%) of 74 specimens, revealing 1 NRBC per 100 or more white blood cells by blood smear examination. The mean percent difference between duplicate automated results was 16.7 compared with 78.1 for the duplicate manual results. There were 9 instances in which the XE-2100 either did not detect the presence of more than 8 NRBCs per 100 white blood cells or generated an automated NRBC% of 18.1 or 18.8 when the smear revealed none. All of these were, however, flagged for smear review.
CONCLUSIONS: Overall correlation between the automated and manual results was excellent. The automated method revealed better precision than the manual method. The number of specimens with false automated results was very small, and all were flagged for verification by a smear review.

Mesh:

Year:  2007        PMID: 17616994     DOI: 10.5858/2007-131-1077-AEOTPO

Source DB:  PubMed          Journal:  Arch Pathol Lab Med        ISSN: 0003-9985            Impact factor:   5.534


  3 in total

1.  Evaluation of nucleated red blood cell count by Sysmex XE-2100 in patients with thalassaemia or sickle cell anaemia and in neonates.

Authors:  Sabrina Buoro; Mauro Vavassori; Silvia Pipitone; Anna Benegiamo; Eleonora Lochis; Sabina Fumagalli; Anna Falanga; Marina Marchetti; Alberto Crippa; Cosimo Ottomano; Giuseppe Lippi
Journal:  Blood Transfus       Date:  2015-01-30       Impact factor: 3.443

2.  Hematologic modifications in natalizumab-treated multiple sclerosis patients: An 18-month longitudinal study.

Authors:  Claire Bridel; Yan Beauverd; Kaveh Samii; Patrice H Lalive
Journal:  Neurol Neuroimmunol Neuroinflamm       Date:  2015-06-18

3.  Investigation of morphological changes for the discrimination of nucleated red blood cells and other leukocytes in Sysmex XN hematology analyzer scattergrams using transmission electron microscopy.

Authors:  Masako Kaido; Yuri Takagi; Mari Kono; Fumie Nakazawa; Shiori Yamamoto; Atsushi Wada; Takashi Morikawa
Journal:  Pract Lab Med       Date:  2017-05-10
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.