Literature DB >> 17613904

Practicing paramedics cannot generate or estimate safe endotracheal tube cuff pressure using standard techniques.

Vivek Parwani1, Robert J Hoffman, Allison Russell, Chetan Bharel, Christine Preblick, In-Hei Hahn.   

Abstract

OBJECTIVES: We sought to determine the ability of paramedics to inflate endotracheal tube cuffs within safe pressure limits as well as to estimate the pressure of previously inflated endotracheal tube cuffs by palpation of the pilot balloon.
METHODS: Using a tracheal simulation model, we conducted a prospective, observational, cross-sectional simulation study of licensed, practicing paramedics. This included evaluation of their ability to inflate the cuff of an endotracheal tube to a safe pressure, defined as < or = 25 cm H(2)O, as well as to identify excessive intracuff pressure in previously inflated ETT cuffs by palpation of the pilot balloon.
RESULTS: Fifty-three paramedics were sampled. The average pressure generated by inflating the endotracheal tube cuff was > 108 cm H(2)O. Participants were only 13% sensitive detecting over inflated endotracheal tube cuffs (95% CI 7.3-17.8).
CONCLUSIONS: Participants were unable to inflate endotracheal tube cuff to safe pressures and were unable to identify endotracheal tube cuffs with excessive intracuff pressure by palpation. Clinicians should consider using devices such as manometers to facilitate safe inflation and accurate measurement of endotracheal tube cuff pressure.

Entities:  

Mesh:

Year:  2007        PMID: 17613904     DOI: 10.1080/10903120701348248

Source DB:  PubMed          Journal:  Prehosp Emerg Care        ISSN: 1090-3127            Impact factor:   3.077


  9 in total

1.  Is sealing cuff pressure, easy, reliable and safe technique for endotracheal tube cuff inflation?: A comparative study.

Authors:  Roshdi R Al-Metwalli; Abdulmohsen A Al-Ghamdi; Hany A Mowafi; Sayed Sadek; Mohammed Abdulshafi; Wesam F Mousa
Journal:  Saudi J Anaesth       Date:  2011-04

2.  Endotracheal Tube Cuff Pressure Assessment: Education May Improve but not Guarantee the Safety of Palpation Technique.

Authors:  Seyed Alireza Seyed Siamdoust; Masood Mohseni; Arash Memarian
Journal:  Anesth Pain Med       Date:  2015-06-22

Review 3.  Mapping the use of simulation in prehospital care - a literature review.

Authors:  Anna Abelsson; Ingrid Rystedt; Björn-Ove Suserud; Lillemor Lindwall
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2014-03-28       Impact factor: 2.953

4.  Endotracheal Tube Cuff Pressures in Patients Intubated Prior to Helicopter EMS Transport.

Authors:  Joseph Tennyson; Tucker Ford-Webb; Stacy Weisberg; Donald LeBlanc
Journal:  West J Emerg Med       Date:  2016-09-13

5.  Does the minimal occlusive volume technique provide adequate endotracheal tube cuff pressure to prevent air leakage?: a prospective, randomized, crossover clinical study.

Authors:  Ha Yeon Park; Mina Kim; Junyong In
Journal:  Anesth Pain Med (Seoul)       Date:  2020-07-31

6.  Linear correlation of endotracheal tube cuff pressure and volume.

Authors:  Robert J Hoffman; Jefrey R Dahlen; Daniela Lipovic; Kai M Stürmann
Journal:  West J Emerg Med       Date:  2009-08

7.  Prospective observational study on tracheal tube cuff pressures in emergency patients--is neglecting the problem the problem?

Authors:  Falko Harm; Mathias Zuercher; Marco Bassi; Wolfgang Ummenhofer
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2013-12-04       Impact factor: 2.953

8.  Endotracheal tube cuff pressures and tube position in critically injured patients on arrival at a referral centre: Avoidable harm?

Authors:  Timothy C Hardcastle; Michael Faurie; David J J Muckart
Journal:  Afr J Emerg Med       Date:  2015-11-21

9.  [Subjective method for tracheal tube cuff inflation: performance of anesthesiology residents and staff anesthesiologists. Prospective observational study].

Authors:  Nadia Maria da Conceição Duarte; Ana Maria Menezes Caetano; Gustavo de Oliveira Arouca; Andrea Tavares Ferrreira; José Luiz de Figueiredo
Journal:  Braz J Anesthesiol       Date:  2020-02-19
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.