CT angiography of renal artery anatomy for evaluating embolic protection devices.

Mallik R Thatipelli1, Enrique A Sabater, Haraldur Bjarnason, Michael A McKusick, Sanjay Misra.   

Abstract

PURPOSE: To determine which commercially available embolic protection devices are suitable for use in renal arteries.
MATERIAL AND METHODS: A retrospective study was conducted of 97 patients who underwent three-dimensional computed tomographic (CT) angiography with either 16- or 64-detector row helical CT of the renal arteries for the evaluation of renal artery stenosis. Nine anatomic measurements were obtained of the renal artery from each reformatted CT angiogram. These data were used to evaluate seven commercially available embolic protection devices (SpideRX, BSC FilterWire EZ, GuardWire, Angioguard, Accunet, Emboshield, TriActiv FX) to determine suitability for use in conjunction with renal artery stent placement. The authors evaluated the reference vessel diameter (diameter to which the stent would be expanded) and length of device.
RESULTS: The mean length (+/- standard deviation) of the main renal artery on the left (39.9 mm +/- 6.7) was significantly less than that on the right (44.9 mm +/- 7.4, P < .001). The mean distal diameter of the main renal artery was significantly larger on the left (6.0 mm +/- 0.5) than on the right (5.6 mm +/- 0.6, P = .02). For a reference vessel diameter of 4-7 mm necessitating a 12-mm-long stent, only three of the seven devices tested (SpideRx, Angioguard, and Accunet) could be used.
CONCLUSION: Three of the seven devices tested could be used for reference vessel diameters of 4-7 mm necessitating a 12-mm-long stent.

Entities:  

Mesh:

Year:  2007        PMID: 17609442     DOI: 10.1016/j.jvir.2007.04.030

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  5 in total

1.  Clinical outcomes in patients with renal artery stenosis treated with stent placement with embolic protection compared with those treated with stent alone.

Authors:  Ankaj Khosla; Sanjay Misra; Eddie L Greene; Axel Pflueger; Steve C Textor; Haraldur Bjarnason; Michael A McKusick
Journal:  Vasc Endovascular Surg       Date:  2012-06-11       Impact factor: 1.089

2.  Embolic protection device use in renal artery stent placement.

Authors:  Mallik R Thatipelli; Sanjay Misra; Srinivas R Sanikommu; Robert M Schainfeld; Sandeep K Sharma; Peter A Soukas
Journal:  J Vasc Interv Radiol       Date:  2009-03-27       Impact factor: 3.464

3.  Variations in Branching Pattern of Renal Artery in Kidney Donors Using CT Angiography.

Authors:  Kumaresan Munnusamy; Sankaran Ponnusamy Kasirajan; Karthikeyan Gurusamy; Gunapriya Raghunath; Shilpakala Leshappa Bolshetty; Sudakshina Chakrabarti; Priyadarshini Annadurai; Zareena Begum Miyajan
Journal:  J Clin Diagn Res       Date:  2016-03-01

4.  Anatomical variations of the renal artery: a computerized tomographic angiogram study in living kidney donors at a Nigerian Kidney Transplant Center.

Authors:  Abayomi Aremu; Martin Igbokwe; Olalekan Olatise; Ahmad Lawal; Kester Maduadi
Journal:  Afr Health Sci       Date:  2021-09       Impact factor: 0.927

Review 5.  Percutaneous revascularization for ischemic nephropathy: the past, present, and future.

Authors:  Stephen C Textor; Sanjay Misra; Gustavo S Oderich
Journal:  Kidney Int       Date:  2012-11-14       Impact factor: 10.612

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.