The use of Speckle Reduction Imaging (SRI) Ultrasound in the characterization of carotid artery plaques.

Nikolaos Liasis1, Chris Klonaris, Athanasios Katsargyris, Sotirios Georgopoulos, Nicos Labropoulos, Chris Tsigris, Athanasios Giannopoulos, Elias Bastounis.   

Abstract

BACKGROUND AND
PURPOSE: Speckle Reduction Imaging is a new algorithm that improves the image quality of B-mode scanning by reducing the reverberation artifacts. In the present study the value of this method for the characterization of atherosclerotic plaques in the internal carotid artery was investigated.
METHODS: Two hundred and twenty two patients (161 men, 61 women; mean age 73 years) referred for carotid ultrasound evaluation were included in the study. Patients with plaques of the internal carotid artery as identified by conventional B-mode scanning were investigated also with the addition of Speckle Reduction Imaging (SRI) with the use of a 4-11-MHz wide band linear transducer. Plaque morphology was rated according to a standardized protocol by two independent observers.
RESULTS: For the determination of plaque echogenicity, the reproducibility of SRI (kappa=0.83) was higher than that of conventional B-mode ultrasound (kappa=0.68). The interobserver agreement for plaque surface characterization was also higher for SRI (kappa=0.8) than for conventional B-mode (kappa=0.61). At the evaluation of the image quality through a semiquantitative analysis, SRI was rated superior in the plaque texture resolution, plaque borders determination, vessel wall demarcation and fibrous cap depiction. In addition, the level of "speckle" was reduced with the use of SRI.
CONCLUSIONS: SRI is a technique that shows good general agreement with high-resolution B-mode and can be used for the characterization of atherosclerotic plaques in the carotid artery. Furthermore, because this advanced technique allows reduction of ultrasound artifacts, it improves the image quality allowing more precise visualization of plaque morphological details.

Entities:  

Mesh:

Year:  2007        PMID: 17604931     DOI: 10.1016/j.ejrad.2007.05.004

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  2 in total

1.  Techniques for Improving Ultrasound Visualization of Biopsy Markers in Axillary Lymph Nodes.

Authors:  Christine Lee; Chenyun Zhou; Brenda Hyde; Pengfei Song; Nicholas Hangiandreou
Journal:  J Clin Imaging Sci       Date:  2020-04-18

2.  Differentiating between adenomyomatosis and gallbladder cancer: revisiting a comparative study of high-resolution ultrasound, multidetector CT, and MR imaging.

Authors:  Sang Heum Bang; Jae Young Lee; Hyunsik Woo; Ijin Joo; Eun Sun Lee; Joon Koo Han; Byung Ihn Choi
Journal:  Korean J Radiol       Date:  2014-03-07       Impact factor: 3.500

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.