| Literature DB >> 17603578 |
Xavier Baillard1, Mathilde Fouché, Rodolphe Le Targat, Philip G Westergaard, Arnaud Lecallier, Yann Le Coq, Giovanni D Rovera, Sebastien Bize, Pierre Lemonde.
Abstract
We report what we believe to be the first accuracy evaluation of an optical lattice clock based on the S01-->P03 transition of an alkaline earth boson, namely, Sr88 atoms. This transition has been enabled by using a static coupling magnetic field. The clock frequency is determined to be 429228066418009(32)Hz. The isotopic shift between Sr87 and Sr88 is 62188135Hz with fractional uncertainty 5x10(-7). We discuss the necessary conditions to reach a clock accuracy of 10(-17) or less by using this scheme.Entities:
Year: 2007 PMID: 17603578 DOI: 10.1364/ol.32.001812