Literature DB >> 17599694

Extubation force: tape versus endotracheal tube holders.

Jestin Carlson1, James Mayrose, Richard Krause, Dietrich Jehle.   

Abstract

STUDY
OBJECTIVE: Tape is the standard method for securing endotracheal tubes to prevent extubation. This study examines the force required to extubate endotracheal tubes from cadavers with either tape or one of 4 commercially available endotracheal tube holders.
METHODS: Newly deceased, unembalmed cadavers were intubated with standard tracheal intubation techniques. The endotracheal tube was secured with either tape or one of 4 commercially available endotracheal tube holders. The endotracheal tube was then connected to a force-measuring device and pulled until the cuff was removed from the trachea. The largest force recorded on the device was then marked as the "extubation force" for that trial.
RESULTS: When tape was used to secure the endotracheal tube, it required a significantly larger force to extubate than 3 of 4 off-the-shelf endotracheal tube holders. Only the Thomas Tube Holder secured the endotracheal tube better than tape.
CONCLUSION: Although the Thomas Tube Holder had the greatest holding force in this study, tape was shown to be the least expensive and outperformed 3 other commercially available devices used to secure endotracheal tubes.

Mesh:

Year:  2007        PMID: 17599694     DOI: 10.1016/j.annemergmed.2007.05.013

Source DB:  PubMed          Journal:  Ann Emerg Med        ISSN: 0196-0644            Impact factor:   5.721


  10 in total

1.  The role of the endotracheal tube cuff in microaspiration.

Authors:  V Anne Hamilton; Mary Jo Grap
Journal:  Heart Lung       Date:  2011-12-30       Impact factor: 2.210

2.  Ivy Loop Wiring: A Useful Form of Endotracheal Tube Stabilization in Burn Patients.

Authors:  Matthew W T Curran; Edward E Tredget
Journal:  Plast Surg (Oakv)       Date:  2017-07-25       Impact factor: 0.947

3.  Two Different Endotracheal Tube Securing Techniques: Fixing Bandage vs. Adhesive Tape.

Authors:  Javad Seyedhosseini; Mojtaba Ahmadi; Amir Nejati; Ali Ardalan; Mohammadhossein Ghafari; Elnaz Vahidi
Journal:  Adv J Emerg Med       Date:  2017-10-13

4.  [Methods for endotracheal tube fixation. Results of a survey of intensive care nurses].

Authors:  O Rothaug; A Müller-Wolff; R Kaltwasser; R Dubb; C Hermes
Journal:  Med Klin Intensivmed Notfmed       Date:  2013-07-14       Impact factor: 0.840

5.  A novel technique for securing supra-glottic airway device to prevent mal-positioning.

Authors:  Arindam Chatterjee; Ashish Kumar Kannaujia
Journal:  Indian J Anaesth       Date:  2016-07

6.  A Comparison of the Haider Tube-Guard® Endotracheal Tube Holder Versus Adhesive Tape to Determine if This Novel Device Can Reduce Endotracheal Tube Movement and Prevent Unplanned Extubation.

Authors:  Jack C Buckley; Adam P Brown; John S Shin; Kirsten M Rogers; Nir N Hoftman
Journal:  Anesth Analg       Date:  2016-05       Impact factor: 5.108

Review 7.  Unplanned Extubations in Intensive Care Unit: evidences for risk factors. A literature review.

Authors:  Chiara Cosentino; Mattia Fama; Chiara Foà; Giorgia Bromuri; Serena Giannini; Marco Saraceno; Angela Spagnoletta; Mbemo Tenkue; Elena Trevisi; Leopoldo Sarli
Journal:  Acta Biomed       Date:  2017-11-30

8.  The effect of adhesive tape versus endotracheal tube fastener in critically ill adults: the endotracheal tube securement (ETTS) randomized controlled trial.

Authors:  Janna S Landsperger; Jesse M Byram; Bradley D Lloyd; Todd W Rice
Journal:  Crit Care       Date:  2019-05-07       Impact factor: 9.097

9.  Extubation force depends upon angle of force application and fixation technique: a study of 7 methods.

Authors:  Jennifer L Wagner; Robin Shandas; Craig J Lanning
Journal:  BMC Anesthesiol       Date:  2014-08-24       Impact factor: 2.217

10.  Real-Time Optical Monitoring of Endotracheal Tube Displacement.

Authors:  Ramzan Ullah; Karl Doerfer; Pawjai Khampang; Faraneh Fathi; Wenzhou Hong; Joseph E Kerschner; Bing Yu
Journal:  Biosensors (Basel)       Date:  2020-11-12
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.