Literature DB >> 17599512

A reinforced sternal wiring technique for transverse thoracosternotomy closure in bilateral lung transplantation: from biomechanical test to clinical application.

Takahiro Oto1, Ramkumar Venkatachalam, Yos S Morsi, Silvana Marasco, Adrian Pick, Marc Rabinov, Franklin Rosenfeldt.   

Abstract

OBJECTIVES: A high incidence of failure of transverse thoracosternotomy closure, involving the loops of wire cutting through the sternum, remains a significant morbidity after bilateral lung transplantation. We postulated that placing peristernal wires inside the usual longitudinal wires could prevent the longitudinal wires from cutting through the sternum. The aims of this study were to investigate the biomechanical and clinical efficacy of the proposed reinforced sternal closure technique.
METHODS: In vitro, 24 artificial sternal models were wired with the reinforced or conventional wiring techniques and were tested either by means of longitudinal distraction or anterior-posterior shear (n = 6 per group). In vivo, the 6-month outcomes of 70 bilateral lung transplantations, including 27 reinforced and 43 conventional wiring techniques, were assessed.
RESULTS: Reinforced wiring was stronger than conventional wiring for both longitudinal distraction (yield load: 585 +/- 60 vs 334 +/- 21 N [P = .03]; maximum load: 807 +/- 60 vs 525 +/- 34 N [P = .03]; postyield stiffness: 91.0 +/- 22.0 vs 32.8 +/- 11.8 N/mm [P = .04]) and anterior-posterior shear (yield load: 405 +/- 9 vs 364 +/- 16 N [P = .03]; postyield stiffness: 47.4 +/- 6.1 vs 27.5 +/- 5.1 N/mm [P = .04]). In multivariate analysis, the use of the conventional wiring technique (odds ratio, 5.38; P = .04) and osteoporosis (odds ratio, 18.31; P = .0005) were significant risk factors associated with sternal dehiscence. In the patients with osteoporosis (n = 25), the incidence of sternal dehiscence in the reinforced wiring group (4/16 [25%]) was significantly lower than that in the conventional wiring group (7/9 [78%], P = .02).
CONCLUSION: Osteoporosis is a significant risk factor for sternal dehiscence after bilateral lung transplantation. The new reinforced sternal wiring technique provides biomechanically superior fixation of the sternum and clinically reduces the incidence of sternal dehiscence in high-risk osteoporotic patients undergoing bilateral lung transplantation.

Entities:  

Mesh:

Year:  2007        PMID: 17599512     DOI: 10.1016/j.jtcvs.2007.03.003

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  4 in total

1.  Clamshell incision versus left anterolateral thoracotomy. Which one is faster when performing a resuscitative thoracotomy? The tortoise and the hare revisited.

Authors:  Alexandros N Flaris; Eric R Simms; Nicolas Prat; Floran Reynard; Jean-Louis Caillot; Eric J Voiglio
Journal:  World J Surg       Date:  2015-05       Impact factor: 3.352

2.  Sternal wound dehiscence after transverse thoracosternotomy for bilateral lung transplantation: report of a case.

Authors:  Fengshi Chen; Akihiro Aoyama; Nobuyuki Kondo; Takuji Fujinaga; Tsuyoshi Shoji; Mitsugu Omasa; Hiroaki Sakai; Toru Bando
Journal:  Surg Today       Date:  2008-09-27       Impact factor: 2.549

3.  Surgical management of extensive dissecting thoracic aortic aneurysm via the semi-clamshell approach.

Authors:  Hiroshi Furukawa; Takeshi Honda; Takahiko Yamasawa; Hisao Masaki; Kazuo Tanemoto
Journal:  Gen Thorac Cardiovasc Surg       Date:  2018-02-22

4.  Experience using a flexible reinforced fiber suture for sternal closure in bilateral lung transplantation recipients undergoing bilateral transverse thoracosternotomy.

Authors:  Jenalee N Coster; Ernest G Chan; Masashi Furukawa; Pablo G Sanchez
Journal:  JTCVS Tech       Date:  2022-05-19
  4 in total