A Snigirev1, A Bjeoumikhov, A Erko, I Snigireva, M Grigoriev, V Yunkin, M Erko, S Bjeoumikhova.
Abstract
A two-step focusing set-up combining a Fresnel zone plate with an ellipsoidal capillary is presented. It is shown that, in addition to the anticipated gain in flux, the employment of the prefocusing micro-optic makes optimal use of the elliptical shape of the capillary by almost eliminating aberrations. A small cross section of the prefocused beam allows a tiny fraction of the capillary surface to be selected, thus reducing the influence of slope errors. An X-ray beam with a 15 keV energy was focused down to a spot size as small as 250 nm, demonstrating the best value that has been achieved up to now for single-bounce capillaries. The use of an ellipsoidal capillary as a micromirror under off-axis illumination by microfocusing optics may open up new opportunities in nanofocusing developments.Year: 2007 PMID: 17587657 DOI: 10.1107/S0909049507025174