OBJECTIVES: The aim of this study was to conduct a field study to survey the performance of digital radiography and how it was used by dentists in general dental practice.
METHODS: 19 general dental practitioners were visited at their clinics. Ambient light (illuminance) was measured in the rooms where the monitors were placed. Different technical display parameters were noted. Test images and two phantoms--one low-contrast phantom and one line-pair resolution phantom--were used to evaluate the digital system. How the dentists used the enhancement program was investigated by noting which functions were used.
RESULTS: Average illuminance in the operating room was 668 lux (range 190-1250 lux). On radiographs of the low-contrast phantom taken at the clinic, the ability to observe the holes decreased as illuminance increased. On average, the "light percentage" initially set on the monitor had to be decreased by 17% and contrast by 10% to optimize the display of the test images. The general dental practitioners used the enhancement programs most often to alter brightness and contrast to obtain the subjectively best image. Large differences between the clinics were noted.
CONCLUSION: Knowledge of how to handle digital equipment in general dental practice should be improved. A calibrated monitor of good quality should be a given priority, as should proper ambient light conditions. There is a need to develop standardized quality controls for digital dental radiography.

Entities:  

Mesh:

Year:  2007        PMID: 17586850     DOI: 10.1259/dmfr/95125494

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  9 in total

1.  Effect of ambient light level at the monitor surface on digital radiographic evaluation of approximal carious lesions: an in vitro study.

Authors:  K Hellén-Halme; A Lith
Journal:  Dentomaxillofac Radiol       Date:  2012-03       Impact factor: 2.419

2.  Effect of display type, DICOM calibration and room illuminance in bitewing radiographs.

Authors:  Soili Kallio-Pulkkinen; Sisko Huumonen; Marianne Haapea; Esa Liukkonen; Annina Sipola; Osmo Tervonen; Miika T Nieminen
Journal:  Dentomaxillofac Radiol       Date:  2015-08-03       Impact factor: 2.419

3.  Carious lesions: diagnostic accuracy using pre-calibrated monitor in various ambient light levels: an in vitro study.

Authors:  K Hellén-Halme; A Lith
Journal:  Dentomaxillofac Radiol       Date:  2013-06-17       Impact factor: 2.419

4.  The usage of digital radiography and cone beam computed tomography among Turkish dentists.

Authors:  Semanur Dölekoğlu; E Fişekçioğlu; M İlgüy; D İlgüy
Journal:  Dentomaxillofac Radiol       Date:  2011-09       Impact factor: 2.419

5.  Comparison between DICOM-calibrated and uncalibrated consumer grade and 6-MP displays under different lighting conditions in panoramic radiography.

Authors:  S Kallio-Pulkkinen; M Haapea; E Liukkonen; S Huumonen; O Tervonen; M T Nieminen
Journal:  Dentomaxillofac Radiol       Date:  2015-01-07       Impact factor: 2.419

6.  The quality of radiographs accompanying endodontic referrals to a health authority clinic.

Authors:  B S Chong; J Miller; S Sidhu
Journal:  Br Dent J       Date:  2015-07-24       Impact factor: 1.626

7.  Effect of two X-ray tube voltages on detection of approximal caries in digital radiographs. An in vitro study.

Authors:  Kristina Hellén-Halme
Journal:  Clin Oral Investig       Date:  2010-02-09       Impact factor: 3.573

8.  A Subjective Assessment of Perceived Clarity of Indirect Digital Images and Processed Digital Images with Conventional Intra-oral Periapical Radiographs.

Authors:  Suchetha N Malleshi; Mahima V G; Anudeepa Raina; Karthikeya Patil
Journal:  J Clin Diagn Res       Date:  2013-08-01

9.  Impact of viewing conditions on the performance assessment of different computer monitors used for dental diagnostics.

Authors:  Thomas Hastie; Sascha Venske-Parker; Johan K M Aps
Journal:  Imaging Sci Dent       Date:  2021-02-09
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.