Resolution of T-wave oversensing with implantable cardioverter defibrillator generator replacement in an adolescent.

Volkan Tuzcu1.   

Abstract

Inappropriate delivery of therapy due to T-wave oversensing is a frequent complication of intracardiac cardioverter defibrillators (ICD). A 16-year-old adolescent with congenital long QT syndrome type 3 underwent single-chamber implantable cardioverter defibrillator (ICD) (EnTrust VR, Medtronic Inc., Minneapolis, MN, USA) implantation.He received two appropriate and four inappropriate shocks. The inappropriate shocks were due to T-wave oversensing. Following the replacement of ICD with a different brand generator (Vitality 2 VR, Boston Scientific Inc., Natick, MA, USA), no more T-wave oversensing is noted. Proper filtering is imperative in ICDs in order to prevent T-wave oversensing-related inappropriate delivery of therapy.

Entities:  

Mesh:

Year:  2007        PMID: 17584279     DOI: 10.1111/j.1540-8159.2007.00785.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

1.  T wave oversensing.

Authors:  Johnson Francis
Journal:  Indian Pacing Electrophysiol J       Date:  2010-06-05

2.  Two cases of inappropriate implantable cardioverter-defibrillator therapies due to T wave oversensing induced by body twisting.

Authors:  Nobuhiro Nishii; Akihito Miyoshi; Motoki Kubo; Hiroyasu Sugiyama; Motomi Tachibana; Koji Nakagawa; Atsuyuki Watanabe; Hiroshi Morita; Hiroshi Ito
Journal:  J Arrhythm       Date:  2016-11-01

3.  Treatment of symptomatic bradycardia due to T-wave oversensing with implantation of a new generator incorporating delayed decay and threshold start sensitization algorithms.

Authors:  Karim Nathan; Omnia Kamel; Daniel Marks; Andrea M Russo
Journal:  HeartRhythm Case Rep       Date:  2020-07-01
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.