Literature DB >> 17571813

Phase aberration correction using ultrasound radiation force and vibrometry optimization.

Matthew W Urban1, Miguel Bernal, James F Greenleaf.   

Abstract

We describe a phase aberration correction method that uses dynamic ultrasound radiation force to harmonically vibrate an object using amplitude modulated continuous wave ultrasound. The phase of each element of an annular array transducer is adjusted to maximize the radiation force and obtain optimal focus of the ultrasound beam. The maximization of the radiation force is performed by monitoring the velocity of scatterers in the focus region. We present theory that shows focal optimization with radiation force has a well-behaved cost function. Experimental validation is shown by correction of manual defocusing of an annular array as well as correcting for a lens-shaped aberrator placed near the transducer. A Doppler laser vibrometer and a pulse-echo Doppler ultrasound method were used to monitor the velocity of a sphere used as a target for the transducer. By maximizing the radiation force-induced vibration of scatterers in the focal region, the resolution of the ultrasound beam can be recovered after aberration defocusing.

Mesh:

Year:  2007        PMID: 17571813     DOI: 10.1109/tuffc.2007.368

Source DB:  PubMed          Journal:  IEEE Trans Ultrason Ferroelectr Freq Control        ISSN: 0885-3010            Impact factor:   2.725


  5 in total

1.  MR-guided adaptive focusing of therapeutic ultrasound beams in the human head.

Authors:  L Marsac; D Chauvet; B Larrat; M Pernot; B Robert; M Fink; A L Boch; J F Aubry; M Tanter
Journal:  Med Phys       Date:  2012-02       Impact factor: 4.071

2.  MR-guided adaptive focusing of ultrasound.

Authors:  Benoît Larrat; Mathieu Pernot; Gabriel Montaldo; Mathias Fink; Mickaël Tanter
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2010-08       Impact factor: 2.725

3.  Energy-based adaptive focusing of waves: application to noninvasive aberration correction of ultrasonic wavefields.

Authors:  Eric Herbert; Mathieu Pernot; Gabriel Montaldo; Mathias Fink; Mickael Tanter
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2009-11       Impact factor: 2.725

4.  Effects of phase aberration on transabdominal focusing for a large aperture, lowf-number histotripsy transducer.

Authors:  Ellen Yeats; Dinank Gupta; Zhen Xu; Timothy L Hall
Journal:  Phys Med Biol       Date:  2022-07-19       Impact factor: 4.174

Review 5.  Production of acoustic radiation force using ultrasound: methods and applications.

Authors:  Matthew W Urban
Journal:  Expert Rev Med Devices       Date:  2018-10-31       Impact factor: 3.166

  5 in total