A C Hudson1, W C Stolte, D W Lindle, R Guillemin.
Abstract
A curved-crystal x-ray emission spectrometer has been designed and built to measure 2-5 keV x-ray fluorescence resulting from a core-level excitation of gas phase species. The spectrometer can rotate 180 degrees, allowing detection of emitted x rays with variable polarization angles, and is capable of collecting spectra over a wide energy range (20 eV wide with 0.5 eV resolution at the Cl K edge) simultaneously. In addition, the entire experimental chamber can be rotated about the incident-radiation axis by nearly 360 degrees while maintaining vacuum, permitting measurements of angular distributions of emitted x rays.Entities:
Mesh:
Year: 2007 PMID: 17552806 DOI: 10.1063/1.2735933
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523